Marcrum Steven C, Kummer Peter, Kreitmayer Christoph, Steffens Thomas
a Department of Otolaryngology , University Hospital Regensburg , Regensburg , Germany .
b Department of Electrical Engineering and Informatics , Technical University of Munich , Munich , Germany , and.
Int J Audiol. 2016;55(6):325-32. doi: 10.3109/14992027.2016.1143979. Epub 2016 Mar 24.
Despite great progress towards optimizing DPOAE primary tone characteristics, factors such as stimulus and intra-subject emission variability have not been addressed. The purpose of this study was to identify optimal primary tone level relationships when these sources of variability were acknowledged, and to identify any influences of test frequency.
Following coupler-based measurements assessing primary tone level stability, two experiments were conducted. In experiment 1, DPOAE test-retest reliability without probe replacement was measured for f2 = 1-6 kHz with L1 = L2 = 65 dB SPL. In experiment 2, optimal L1-L2 relationships were identified for f2 = 1-6 kHz. For 20 ≤ L2 ≤ 75 dB SPL, L1 was varied 15 dB SPL above and below the recommendation of L1 = 0.4 L2 + 39 [dB SPL].
Eleven normal-hearing adults participated in experiment 1. Thirty normal-hearing adults participated in experiment 2.
Stimulus variability did not exceed 0.1 dB SPL. DPOAE reliability testing revealed an across-frequency mean standard error of measurement of 0.52 dB SPL. The average optimal L1-L2 relationship was described by L1 = 0.49 L2 + 41 [dB SPL]. A significant effect of frequency was identified for 6 kHz.
Including relevant sources of variability improves internal validity of a primary tone level optimization formula.
尽管在优化畸变产物耳声发射(DPOAE)基频特性方面取得了巨大进展,但诸如刺激和个体内发射变异性等因素尚未得到解决。本研究的目的是在考虑这些变异性来源的情况下确定最佳基频水平关系,并确定测试频率的任何影响。
在基于耦合器的测量评估基频水平稳定性之后,进行了两项实验。在实验1中,在L1 = L2 = 65 dB SPL的条件下,测量了f2 = 1 - 6 kHz时不更换探头的DPOAE重测信度。在实验2中,确定了f2 = 1 - 6 kHz时的最佳L1 - L2关系。对于20≤L2≤75 dB SPL,L1在L1 = 0.4L2 + 39 [dB SPL]的推荐值上下15 dB SPL范围内变化。
11名听力正常的成年人参与了实验1。30名听力正常的成年人参与了实验2。
刺激变异性不超过0.1 dB SPL。DPOAE信度测试显示全频率平均测量标准误差为0.52 dB SPL。平均最佳L1 - L2关系由L1 = 0.49L2 + 41 [dB SPL]描述。在6 kHz时发现了频率的显著影响。
考虑相关变异性来源可提高基频水平优化公式的内部效度。