数字减影血管造影中手动与参考点叠加对图像质量的影响。

Effect of manual compared with reference point superimposition on image quality in digital subtraction radiography.

作者信息

Wenzel A

出版信息

Dentomaxillofac Radiol. 1989 Nov;18(4):145-50. doi: 10.1259/dmfr.18.4.2701173.

Abstract

The aim of this study was to compare a new subtraction program based on positioning of reference points (RP) in the two images with the classic manual (M) superimposition of the images during recording. The experiments were performed on 22 dry mandibles from domestic pigs. A bone chip was prepared from the cortex with seven edges: 0.26, 0.30, 0.36, 0.42, 0.49, 0.55 and 0.72 mm thick respectively. Each mandible was radiographed at 0 degrees vertical angulation without the bone chip. The bone chip was then fixed to the lingual plate for a series of radiographs during which the vertical angulation was varied from 0 degrees to 10 degrees at 1 degree intervals. After completion of this series of exposures, amalgam fillings were placed in the two molars on the right side and a second series obtained. The radiographs were digitized (512 x 512 x 8 bit resolution), stored in a personal computer and then subtracted in a random order by the two methods. The RP method was found to be superior to the M method for all the angulations evaluated; as the images were more homogeneous as shown by the smaller SD in the grey-scale histogram. The best results were obtained on images with amalgam-filled teeth. Significantly more bone edges were seen with the RP method, especially when amalgam fillings were present. The RP method may therefore be superior to the M in clinical trials using subtraction radiography.

摘要

本研究的目的是比较一种基于两张图像中参考点(RP)定位的新型减法程序与记录过程中图像的经典手动(M)叠加。实验在22个家猪的干燥下颌骨上进行。从皮质制备了一片骨片,有七条边缘,厚度分别为0.26、0.30、0.36、0.42、0.49、0.55和0.72毫米。每个下颌骨在垂直角度为0度且无骨片的情况下进行放射照相。然后将骨片固定在舌板上进行一系列放射照相,在此期间垂直角度以1度的间隔从0度变化到10度。完成这一系列曝光后,在右侧的两颗磨牙中放置汞合金填充物并获得第二组图像。将放射照片数字化(512×512×8位分辨率),存储在个人计算机中,然后用两种方法以随机顺序进行减法运算。发现在所有评估的角度下,RP方法均优于M方法;如灰度直方图中较小的标准差所示,图像更加均匀。在有汞合金填充牙齿的图像上获得了最佳结果。使用RP方法能看到明显更多的骨边缘,尤其是当存在汞合金填充物时。因此,在使用减法放射照相的临床试验中,RP方法可能优于M方法。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验