使用高精度体外评估方法,不同锥束计算机断层扫描和多层螺旋计算机断层扫描参数对虚拟三维模型准确性的影响。

The impact of different cone beam computed tomography and multi-slice computed tomography scan parameters on virtual three-dimensional model accuracy using a highly precise ex vivo evaluation method.

作者信息

Matta Ragai-Edward, von Wilmowsky Cornelius, Neuhuber Winfried, Lell Michael, Neukam Friedrich W, Adler Werner, Wichmann Manfred, Bergauer Bastian

机构信息

Department of Prosthodontics (Head: Prof. Dr. M. Wichmann), Erlangen University Hospital, Glückstrasse 11, 91054 Erlangen, Germany.

Department of Oral and Maxillofacial Surgery (Head: Prof. Dr. Dr. Dr. h.c. F.W. Neukam), Erlangen University Hospital, Glückstrasse 11, 91054 Erlangen, Germany.

出版信息

J Craniomaxillofac Surg. 2016 May;44(5):632-6. doi: 10.1016/j.jcms.2016.02.005. Epub 2016 Feb 13.

Abstract

OBJECTIVES

Multi-slice computed tomography (MSCT) and cone beam computed tomography (CBCT) are indispensable imaging techniques in advanced medicine. The possibility of creating virtual and corporal three-dimensional (3D) models enables detailed planning in craniofacial and oral surgery. The objective of this study was to evaluate the impact of different scan protocols for CBCT and MSCT on virtual 3D model accuracy using a software-based evaluation method that excludes human measurement errors.

MATERIAL AND METHODS

MSCT and CBCT scans with different manufacturers' predefined scan protocols were obtained from a human lower jaw and were superimposed with a master model generated by an optical scan of an industrial noncontact scanner. To determine the accuracy, the mean and standard deviations were calculated, and t-tests were used for comparisons between the different settings.

RESULTS

Averaged over 10 repeated X-ray scans per method and 19 measurement points per scan (n = 190), it was found that the MSCT scan protocol 140 kV delivered the most accurate virtual 3D model, with a mean deviation of 0.106 mm compared to the master model. Only the CBCT scans with 0.2-voxel resolution delivered a similar accurate 3D model (mean deviation 0.119 mm).

CONCLUSION

Within the limitations of this study, it was demonstrated that the accuracy of a 3D model of the lower jaw depends on the protocol used for MSCT and CBCT scans.

摘要

目的

多层螺旋计算机断层扫描(MSCT)和锥束计算机断层扫描(CBCT)是现代医学中不可或缺的成像技术。创建虚拟和实体三维(3D)模型的可能性使得颅面和口腔外科手术能够进行详细规划。本研究的目的是使用一种基于软件的评估方法,排除人为测量误差,评估CBCT和MSCT不同扫描协议对虚拟3D模型准确性的影响。

材料与方法

从人类下颌骨获取不同制造商预定义扫描协议的MSCT和CBCT扫描图像,并与工业非接触式扫描仪光学扫描生成的主模型进行叠加。为了确定准确性,计算了平均值和标准差,并使用t检验对不同设置进行比较。

结果

每种方法平均进行10次重复X射线扫描,每次扫描有19个测量点(n = 190),结果发现MSCT扫描协议140 kV产生的虚拟3D模型最准确,与主模型相比平均偏差为0.106 mm。只有0.2体素分辨率的CBCT扫描产生了类似准确的3D模型(平均偏差0.119 mm)。

结论

在本研究的局限性内,证明了下颌骨3D模型的准确性取决于用于MSCT和CBCT扫描的协议。