Davies Jessica, Denyer Tom, Hadfield James
Genomics Core Facility, Cancer Research UK Cambridge Institute, Li Ka Shing Centre, University of Cambridge, Robinson Way, Cambridge, UK.
Biotechniques. 2016 Apr 1;60(4):197-9. doi: 10.2144/000114403. eCollection 2016 Apr.
The Agilent 2100 Bioanalyzer enables small-scale gel electrophoretic separation of nucleic acids on a microfluidic chip; however, a shortage of chips and an excess of reagents are common issues. Here we explore the compatibility of two commonly used Bioanalyzer reagents with three Bioanalyzer chip types. Microfluidic electrophoretic separation of DNA and RNA using DNA High Sensitivity and RNA 6000 Nano reagents, respectively, was successfully performed on multiple chip types following the assay-specific protocols. For RNA quality and next-generation sequencing (NGS) library length estimation, the Bioanalyzer chips we tested can be used interchangeably. These findings will be valuable for any laboratory using the Agilent Bioanalyzer in a shared facility.
安捷伦2100生物分析仪能够在微流控芯片上对核酸进行小规模凝胶电泳分离;然而,芯片短缺和试剂过剩是常见问题。在此,我们探究了两种常用的生物分析仪试剂与三种生物分析仪芯片类型的兼容性。按照特定检测方案,分别使用DNA高灵敏度试剂和RNA 6000纳米试剂在多种芯片类型上成功进行了DNA和RNA的微流控电泳分离。对于RNA质量和下一代测序(NGS)文库长度估计,我们测试的生物分析仪芯片可以互换使用。这些发现对于在共享设施中使用安捷伦生物分析仪的任何实验室都将具有重要价值。