Abstract

The purpose of this study was to validate the dosimetric performance of Varian surface applicators with the source vertically positioned and develop procedures for clinical implementation. The Varian surface applicators with the source vertically positioned provide a wide range of apertures making them clinically advantageous, though the steep dose gradient in the region of 3-4 mm prescription depth presents multiple challenges. The following commissioning tests were performed: 1) verification of functional integrity and physical dimensions; and 2) dosimetric measurements to validate data provided by Varian as well as data obtained using the Acuros algorithm for heterogeneity corrected dose calculation. A solid water (SW) phantom was scanned and the Acuros algorithm was used to compute the dose at 5 mm depth and at surface for all applicators. Two sets of reference dose measurements were performed, with the source positioned at (i) -10 mm and (ii) -15 mm from the center of the first nominal dwell position. Measurements were taken at 5 mm depth in a SW phantom and in air at the applicator surface. The results were then compared to the vendor's data and to the Acuros calculated dose. Relative dose measurements using Gafchromic films were taken at a depth of 4 mm in SW. Percent depth ionization (PDI) measurements using ion chamber were performed in SW. The profiles generated from film measurements and the PDI plots were compared with those computed using the Acuros algorithm and vendor's data, when available. Preliminary leakage tests were performed using optically stimulated luminescence dosimeters (OSLDs) and the results were compared with Acuros predictions. All applicators were found to be functional with physical dimensions within 1 mm of specifications. For scenario (ii) measurements taken in SW at 5 mm depth and in air at the surface of each applicator were within 10% and 4% agreement with vendor's data, respectively. Compared with Acuros predictions, these measurements were within 6% and 5%, respectively. Measurements taken for scenario (i) showed reduced agreement with both the vendor's data as well as the Acuros calculations, especially when using the 10 mm applicator. The full widths of the measured dose profiles were within 2 mm of those predicted by Acuros at the 90% dose level. The PDI plots and measured leakage dose were in good agreement with vendor's data and Acuros predictions. Based on the dosimetric results, a quality assurance program and procedures for clinical implementation were developed. Treatment planning will be performed using scenario (ii). The 10mm applicator will not be released for clinical use. A prescription depth of 4mm is recommended, to ensure full coverage at 3 mm and a minimum dose of 90% of prescribed dose at 4 mm depth.

摘要

本研究的目的是验证瓦里安表面敷贴器在源垂直放置时的剂量学性能,并制定临床实施程序。源垂直放置的瓦里安表面敷贴器提供了广泛的孔径,使其在临床上具有优势,尽管在3 - 4毫米处方深度区域的陡峭剂量梯度带来了多重挑战。进行了以下调试测试:1)功能完整性和物理尺寸的验证;2)剂量学测量,以验证瓦里安提供的数据以及使用Acuros算法进行不均匀性校正剂量计算所获得的数据。对固体水(SW)模体进行扫描,并使用Acuros算法计算所有敷贴器在5毫米深度和表面的剂量。进行了两组参考剂量测量,源分别位于(i)距第一个标称驻留位置中心-10毫米和(ii)-15毫米处。在SW模体中5毫米深度处以及敷贴器表面的空气中进行测量。然后将结果与供应商的数据以及Acuros计算的剂量进行比较。使用Gafchromic胶片在SW中4毫米深度处进行相对剂量测量。使用电离室在SW中进行百分深度电离(PDI)测量。将胶片测量生成的数据曲线和PDI图与使用Acuros算法和供应商数据(如可用)计算得到的结果进行比较。使用光激励发光剂量计(OSLD)进行初步泄漏测试,并将结果与Acuros预测值进行比较。发现所有敷贴器功能正常,物理尺寸在规格的1毫米范围内。对于情况(ii),在SW中5毫米深度处以及每个敷贴器表面空气中的测量结果与供应商数据的一致性分别在10%和4%以内。与Acuros预测值相比,这些测量结果分别在6%和5%以内。情况(i)的测量结果与供应商数据以及Acuros计算结果的一致性较低,尤其是使用10毫米敷贴器时。在90%剂量水平下,测量的剂量曲线半高宽与Acuros预测值相差在2毫米以内。PDI图和测量的泄漏剂量与供应商数据和Acuros预测值吻合良好。基于剂量学结果,制定了质量保证计划和临床实施程序。治疗计划将采用情况(ii)。10毫米敷贴器将不发放用于临床。建议处方深度为4毫米,以确保在3毫米处完全覆盖,并且在4毫米深度处的最小剂量为处方剂量的90%。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/467e/5874946/8c1ef5d9745e/ACM2-17-231-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验