Stokes D C, Rao B N, Mirro J, Mackert P W, Austin B, Colten M, Hancock M L
Department of Surgery, St Jude Children's Research Hospital, Memphis, TN 38101.
J Pediatr Surg. 1989 Mar;24(3):257-62. doi: 10.1016/s0022-3468(89)80007-7.
Thrombotic occlusion of Hickman and Broviac central venous catheters is a serious obstacle to their long-term use. Because resistance to flow (R) through a catheter of lumen radius, r, is proportional to 1/r4, we hypothesized that measurement of R would provide an objective and sensitive monitor for partial occlusions. Our measurements showed that median R at a flow of 17 mL/min was 0.7 cmH2O/mL/min in normally functioning Hickman catheters, and 4.1 cmH2O/mL/min in Broviac catheters. In obstructed catheters, which by subjective standards resisted flushing or blood withdrawal, median R was 3.0 cmH2O/mL/min for Hickman and 5.6 cmH2O/mL/min for Broviac catheters, representing significant increases. In a series of obstructed lines in which urokinase was administered, R decreased from 7.7 to 4.5 in Hickman catheters and from 5.6 to 4.2 in obstructed Broviac catheters. The elevated resistance in Hickman catheters after urokinase suggested that residual catheter obstruction was present even though catheter function returned to normal. Elevated R was seen with abnormal venograms in seven of 13 patients. Four patients had normal R values and abnormal venograms, and two patients had elevated R values with normal venograms. Measurement of resistance in Hickman and Broviac catheters provides a simple technique that can supplement or replace venography in the serial assessment and treatment of partial obstruction.
希克曼(Hickman)和布罗维阿克(Broviac)中心静脉导管的血栓形成性闭塞是其长期使用的严重障碍。由于通过内径为r的导管的血流阻力(R)与1/r⁴成正比,我们推测测量R可为部分闭塞提供客观且灵敏的监测指标。我们的测量结果显示,在正常功能的希克曼导管中,流速为17 mL/min时的R中位数为0.7 cmH₂O/mL/min,而在布罗维阿克导管中为4.1 cmH₂O/mL/min。在主观判断难以冲洗或抽血的阻塞导管中,希克曼导管的R中位数为3.0 cmH₂O/mL/min,布罗维阿克导管为5.6 cmH₂O/mL/min,均显著升高。在一系列接受尿激酶治疗的阻塞导管中,希克曼导管的R从7.7降至4.5,阻塞的布罗维阿克导管的R从5.6降至4.2。尿激酶治疗后希克曼导管的阻力升高表明,尽管导管功能恢复正常,但仍存在残留的导管阻塞。13例患者中有7例静脉造影异常时出现R升高。4例患者R值正常但静脉造影异常,2例患者R值升高但静脉造影正常。测量希克曼和布罗维阿克导管的阻力提供了一种简单的技术,可在部分阻塞的连续评估和治疗中补充或替代静脉造影。