Phys Med. 2016 May;32(5):681-90. doi: 10.1016/j.ejmp.2016.04.011. Epub 2016 May 6.
Within the SYRMA-CT collaboration based at the ELETTRA synchrotron radiation (SR) facility the authors investigated the imaging performance of the phase-contrast computed tomography (CT) system dedicated to monochromatic in vivo 3D imaging of the female breast, for breast cancer diagnosis.
Test objects were imaged at 38keV using monochromatic SR and a high-resolution CdTe photon-counting detector. Signal and noise performance were evaluated using modulation transfer function (MTF) and noise power spectrum. The analysis was performed on the images obtained with the application of a phase retrieval algorithm as well as on those obtained without phase retrieval. The contrast to noise ratio (CNR) and the capability of detecting test microcalcification clusters and soft masses were investigated.
For a voxel size of (60μm)(3), images without phase retrieval showed higher spatial resolution (6.7mm(-1) at 10% MTF) than corresponding images with phase retrieval (2.5mm(-1)). Phase retrieval produced a reduction of the noise level and an increase of the CNR by more than one order of magnitude, compared to raw phase-contrast images. Microcalcifications with a diameter down to 130μm could be detected in both types of images.
The investigation on test objects indicates that breast CT with a monochromatic SR source is technically feasible in terms of spatial resolution, image noise and contrast, for in vivo 3D imaging with a dose comparable to that of two-view mammography. Images obtained with the phase retrieval algorithm showed the best performance in the trade-off between spatial resolution and image noise.
在 ELETTRA 同步加速器辐射(SR)设施的 SYRMA-CT 合作项目中,作者研究了专门用于女性乳房单色体内 3D 成像的相位对比计算机断层扫描(CT)系统的成像性能,用于乳腺癌诊断。
使用单色 SR 和高分辨率 CdTe 光子计数探测器对测试物体进行 38keV 成像。使用调制传递函数(MTF)和噪声功率谱评估信号和噪声性能。对应用相位恢复算法获得的图像以及未应用相位恢复算法获得的图像进行了分析。研究了对比度噪声比(CNR)以及检测测试微钙化簇和软组织的能力。
对于体素尺寸为(60μm)(3),未进行相位恢复的图像显示出比具有相位恢复的图像更高的空间分辨率(10% MTF 时为 6.7mm(-1))。与原始相衬图像相比,相位恢复降低了噪声水平并将 CNR 提高了一个数量级以上。在两种类型的图像中都可以检测到直径达 130μm 的微钙化。
对测试物体的研究表明,使用单色 SR 源的乳腺 CT 在空间分辨率、图像噪声和对比度方面在技术上是可行的,适用于与双视图乳房 X 光摄影相比剂量相当的体内 3D 成像。使用相位恢复算法获得的图像在空间分辨率和图像噪声之间的折衷中表现出最佳性能。