Radiol Oncol. 2018 Mar 24;52(3):329-336. doi: 10.2478/raon-2018-0015.
Background The SYRMA-3D collaboration is setting up the first clinical trial of phase-contrast breast CT with synchrotron radiation at the Elettra synchrotron facility in Trieste, Italy. In this communication, a quality control protocol for breast CT is proposed, and a first test of image quality measurements is performed by means of a custom-made radiographic phantom. Materials and methods A set of projections is acquired and used to perform a CT reconstruction of two selected portions of the phantom. Such portions contain a uniform layer of water and a set of radiographic inserts, respectively. Together, they allow to perform several image quality measurements, namely CT number linearity, reconstruction accuracy, uniformity, noise, and low contrast resolution. All measurements are repeated at different beam energies in the range of interest, and at two different dose values. Results Measurements show a good linearity in the soft tissue range, paired to a high accuracy of the CT number reconstruction. Uniformity and noise measurements show that reconstruction inhomogeneities are bound to a few percent of the average pixel values. However, low contrast detectability is limited to the higher portion of the explored energy range. Conclusions The results of the measurements are satisfactory in terms of their quality, feasibility and reproducibility. With minimal modifications, the phantom is promising to allow a set of image quality measurements to be used in the upcoming clinical trial.
SYRMA-3D 合作组织正在意大利的的里雅斯特 Elettra 同步加速器设施中开展首例利用同步加速器辐射进行对比增强乳腺 CT 的临床试验。在本通讯中,提出了一种乳腺 CT 的质量控制方案,并通过定制的射线照相体模进行了图像质量测量的初步测试。
采集一组投影,并用于对体模的两个选定部分进行 CT 重建。这些部分分别包含一层均匀的水层和一组射线照相插件。它们共同允许进行多项图像质量测量,即 CT 值线性度、重建准确性、均匀性、噪声和低对比度分辨率。所有测量都在感兴趣的范围内的不同光束能量和两个不同剂量值下重复进行。
测量结果显示软组织范围内具有良好的线性度,同时 CT 值重建具有很高的准确性。均匀性和噪声测量结果表明,重建不均匀性仅限于平均像素值的几个百分点。然而,低对比度检测能力仅限于所探索的能量范围的较高部分。
就其质量、可行性和可重复性而言,测量结果令人满意。稍加修改,该体模有望允许在即将进行的临床试验中使用一组图像质量测量。