Evaluation of two commercially available ELISA kits for the determination of melatonin concentrations in amniotic fluid throughout pregnancy.

作者信息

Bagci Soyhan, Altuntas Özlem, Katzer David, Berg Christoph, Willruth Arne, Reutter Heiko, Bartmann Peter, Müller Andreas, Zur Berndt

机构信息

1 Neonatology and Pediatric Intensive Care, Children's Hospital, University of Bonn, Bonn, Germany.

2 Department of Obstetrics and Prenatal Medicine, University of Bonn, Bonn, Germany.

出版信息

Ann Clin Biochem. 2017 Jan;54(1):107-112. doi: 10.1177/0004563216645123. Epub 2016 Sep 28.

Abstract

Background The aim of the present study is to evaluate the utility of extraction versus non-extraction-based commercial melatonin ELISA kits for determining the melatonin concentration in amniotic fluid obtained in early and late pregnancy. Methods Pregnancy duration less than 28 weeks was defined as early and from 28 weeks until delivery as late gestation. Nine samples were obtained in early and 18 in late pregnancy. Two commercially available melatonin ELISA kits (melatonin ELISA RE54021, including methanol-based extraction and direct saliva melatonin ELISA RE 54041, not including an extraction step, both from IBL-International, Germany) were used to determine melatonin concentrations in amniotic fluid. Results The mean melatonin concentration in ELISAs assayed by the non-extraction was significantly lower than those assayed after extraction. Subgroup analysis showed that there was no significant difference between melatonin concentration measured by non-extraction versus extraction ELISA in early pregnancy (11.2 ± 7.4 vs. 12.2 ± 7.7, respectively, P = 0.463) but that the mean melatonin concentration in late pregnancy was significantly lower when assayed by non-extraction ELISA than when assayed by extraction ELISA (14.8 ± 9.3 vs. 145.1 ± 179.3, respectively; P < 0.001). Agreement between both measurements in late pregnancy was rather poor (r= 0.271, P = 0.022), as opposed to the good correlation found in early pregnancy (r= 0.929, P < 0.001). Conclusions The present study revealed that a melatonin assay without an extraction step, such as direct saliva ELISA, does not seem to be a valid method to determine the melatonin concentration of amniotic fluid, especially in late gestation.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验