Suppr超能文献

Stability-indicating high-performance liquid chromatographic assay for the determination of metoclopramide hydrochloride in pharmaceutical dosage forms.

作者信息

Suleiman M S, Najib N M, el-Sayed Y M, Badwan A

出版信息

Analyst. 1989 Mar;114(3):365-8. doi: 10.1039/an9891400365.

Abstract

A simple, reliable and selective high-performance liquid chromatographic method for the determination of metoclopramide hydrochloride in pharmaceutical dosage forms has been developed and evaluated. The drug and the internal standard (phenobarbitone) were eluted from a 5-micron C8 reversed-phase column at ambient temperature with a mobile phase consisting of phosphate buffer (10 mM)-methanol-acetonitrile (50 + 28 + 22) adjusted to pH 4.8 with orthophosphoric acid. The mobile phase was pumped at a flow-rate of 1.5 ml min-1 and the effluent was monitored spectrophotometrically at 214 nm. The retention times of the internal standard and metoclopramide hydrochloride were 3.0 and 7.5 min, respectively. Quantification was achieved by measuring the peak-height ratio of the drug to the internal standard. A linear relationship was found over the range 1-10 micrograms ml-1. Within-day coefficients of variation (CVs) ranged from 0.50 to 1.70% and between-day CVs from 0.68 to 4.07% at three different concentrations. The developed procedure was compared with the current BP method for the assay of metoclopramide hydrochloride in tablets. The proposed method was also used to study the stability of metoclopramide hydrochloride.

摘要