Goel V K, Clark C R, Harris K G, Kim Y E, Schulte K R
Department of Biomedical Engineering, University of Iowa, Iowa City 52242.
Ann Biomed Eng. 1989;17(2):115-26. doi: 10.1007/BF02368022.
The effects of facet wiring procedure commonly used for stabilizing cervical spines after laminectomy or bilateral facet dislocation on the motion behavior of whole cervical spines are investigated using a Selspot II system. A fresh human ligamentous intact specimen was potted at T1/T2 vertebra and clinically relevant loads applied to the topmost vertebra (C2) of the specimen. The resulting three rotational components of each of the five vertebral bodies (C3-C7) were recorded. Specimen was injured to mimic total laminectomies at C5 and C6 vertebral levels and tested again. The injured specimen was stabilized, using a facet wiring construct, across C4-C7 segment before testing for the final time. The injured specimens, compared to the intact specimens, demonstrated an increase in flexion-extension of about 10%. Facet wiring imparted stability to the cervical spine by stiffening segments up to roughly four times intact values.