Takegami Kazuki, Hayashi Hiroaki, Yamada Kenji, Mihara Yoshiki, Kimoto Natsumi, Kanazawa Yuki, Higashino Kousaku, Yamashita Kazuta, Hayashi Fumio, Okazaki Tohru, Hashizume Takuya, Kobayashi Ikuo
Graduate School of Health Sciences, Tokushima University, 3-18-5 Kuramoto-cho, Tokushima, Tokushima, 770-8503, Japan.
Yamaguchi University Hospital, 1-1-1, Minamikogushi, Ube, Yamaguchi, 755-8505, Japan.
Radiol Phys Technol. 2017 Mar;10(1):49-59. doi: 10.1007/s12194-016-0366-1. Epub 2016 Jun 24.
Entrance surface dose (ESD) measurements are important in X-ray computed tomography (CT) for examination, but in clinical settings it is difficult to measure ESDs because of a lack of suitable dosimeters. We focus on the capability of a small optically stimulated luminescence (OSL) dosimeter. The aim of this study is to propose a practical method for using an OSL dosimeter to measure the ESD when performing a CT examination. The small OSL dosimeter has an outer width of 10 mm; it is assumed that a partial dose may be measured because the slice thickness and helical pitch can be set to various values. To verify our method, we used a CT scanner having 320 rows of detectors and checked the consistencies of the ESDs measured using OSL dosimeters by comparing them with those measured using Gafchromic™ films. The films were calibrated using an ionization chamber on the basis of half-value layer estimation. On the other hand, the OSL dosimeter was appropriately calibrated using a practical calibration curve previously proposed by our group. The ESDs measured using the OSL dosimeters were in good agreement with the reference ESDs from the Gafchromic™ films. Using these data, we also estimated the uncertainty of ESDs measured with small OSL dosimeters. We concluded that a small OSL dosimeter can be considered suitable for measuring the ESD with an uncertainty of 30 % during CT examinations in which pitch factors below 1.000 are applied.
在X射线计算机断层扫描(CT)检查中,体表入射剂量(ESD)测量很重要,但在临床环境中,由于缺乏合适的剂量计,很难测量ESD。我们关注小型光激励发光(OSL)剂量计的性能。本研究的目的是提出一种在进行CT检查时使用OSL剂量计测量ESD的实用方法。小型OSL剂量计的外部宽度为10毫米;由于切片厚度和螺旋螺距可以设置为各种值,因此可以测量部分剂量。为了验证我们的方法,我们使用了一台具有320排探测器的CT扫描仪,并通过将使用OSL剂量计测量的ESD与使用Gafchromic™ 胶片测量的ESD进行比较,检查了测量结果的一致性。这些胶片基于半值层估计使用电离室进行校准。另一方面,OSL剂量计使用我们小组先前提出的实用校准曲线进行了适当校准。使用OSL剂量计测量的ESD与Gafchromic™ 胶片的参考ESD高度一致。利用这些数据,我们还估计了使用小型OSL剂量计测量ESD的不确定度。我们得出结论,在应用螺距因子低于1.000的CT检查中,小型OSL剂量计可被认为适用于测量ESD,不确定度为30%。