Lindner Thomas, Larsen Naomi, Jansen Olav, Helle Michael
University Hospital Schleswig-Holstein, Campus Kiel, Clinic for Radiology and Neuroradiology, Arnold-Heller-Straße 3, 24105, Kiel, Germany.
Philips GmbH Innovative Technologies, Research Lab, Röntgenstraße 24-26, 22335, Hamburg, Germany.
MAGMA. 2016 Dec;29(6):843-852. doi: 10.1007/s10334-016-0574-z. Epub 2016 Jun 29.
To accelerate super-selective arterial spin labeling (ASL) angiography by using a single control condition denoted as cycled super-selective arterial spin labeling.
A single non-selective control image is acquired that is shared by selective label images. Artery-selective imaging is possible by geometrically changing the position of the labeling focus to more than one artery of interest during measurement. The presented approach is compared to conventional super-selective imaging in terms of its labeling efficiency inside and outside the labeling focus using numerical simulations and in vivo measurements. Additionally, the signal-to-noise ratios of the images are compared to non-selective ASL angiography and analyzed using a two-way ANOVA test and calculating the Pearson's correlation coefficients.
The results indicate that the labeling efficiency is not reduced within the labeled artery, but can increase as a function of distance to the artery of interest when compared to conventional super-selective ASL. In the final images, no statistically significant difference of image quality can be observed while the acquisition duration could be reduced when the major brain feeding arteries are being tagged.
Using super-selective arterial spin labeling, a single non-selective control acquisition suffices for reconstructing selective angiograms of the cerebral vasculature, thereby accelerating image acquisition of the major intracranial arteries without notable loss of information.
通过使用一种称为循环超选择性动脉自旋标记的单一控制条件来加速超选择性动脉自旋标记(ASL)血管造影。
采集一幅由选择性标记图像共享的单一非选择性对照图像。在测量过程中,通过几何方式将标记焦点的位置改变到多个感兴趣的动脉,从而实现动脉选择性成像。使用数值模拟和体内测量,将所提出的方法与传统超选择性成像在标记焦点内外的标记效率方面进行比较。此外,将图像的信噪比与非选择性ASL血管造影进行比较,并使用双向方差分析测试和计算皮尔逊相关系数进行分析。
结果表明,与传统超选择性ASL相比,标记动脉内的标记效率没有降低,而是可以随着与感兴趣动脉距离的增加而提高。在最终图像中,当主要脑供血动脉被标记时,图像质量没有观察到统计学上的显著差异,而采集持续时间可以缩短。
使用超选择性动脉自旋标记,单次非选择性对照采集足以重建脑血管系统的选择性血管造影,从而加速主要颅内动脉的图像采集,而不会有明显的信息损失。