Sudden Appearance of Water in Flowmeter During Air/Oxygen and Sevoflurane Anaesthesia.

作者信息

Kandemir Tünay, Muslu Selda, Kandemir Erbin

机构信息

Clinic of Anaesthesia, Dr. Abdurrahman Yurtaslan Oncology Training and Research Hospital, Ankara, Turkey.

出版信息

Turk J Anaesthesiol Reanim. 2015 Feb;43(1):65-7. doi: 10.5152/TJAR.2014.57070. Epub 2014 Sep 9.

Abstract

Endotracheal intubation was performed, and a water bubbling sound was heard from the anaesthesia device immediately after the release of gases to administer the O2-air-sevoflurane mixture. The flowmeter on the anaesthesia device was then found to be filled with water. The breakdown of the dryer in the medical air compressor system was determined as the source of the problem, since a greasy fluid mixture was released from the air-wall outlets in all rooms. Consequently, the anaesthesia team should keep in mind that problems as seen in the current case might emerge and should be alert.

摘要

进行了气管插管,在释放气体以输送氧气-空气-七氟醚混合气体后,立即从麻醉设备中听到水泡声。随后发现麻醉设备上的流量计充满了水。由于所有房间的空气壁出口都释放出油腻的液体混合物,确定医用空气压缩机系统中的干燥器故障是问题的根源。因此,麻醉团队应牢记当前病例中出现的问题可能会出现,并应保持警惕。