Abstract

BACKGROUND

Loosening of the glenoid component continues to be the main cause of medium and long-term failure of shoulder replacements. The purpose of this study was to evaluate the early clinical and radiographic results following use of an all-polyethylene pegged glenoid component designed for hybrid (biological and cement) fixation.

METHODS

Eighty-three shoulders in 77 patients (mean age, 68.6 years) underwent total shoulder arthroplasty with a pegged hybrid-fixation component (bone-ingrowth glenoid and cemented peg). Outcomes were determined with the American Shoulder and Elbow Surgeons (ASES) score and the Oxford shoulder score. A detailed analysis of radiographs and fine-slice computed tomography (CT) images was performed to determine the extent of bone ingrowth between the flanges and the extent of radiolucent lines at the prosthesis-bone interface.

RESULTS

The mean duration of follow-up was 46.7 months (range, 24 to 99 months). At the time of final follow-up, the median ASES score was 97 points (range, 43 to 100 points) and the median Oxford score was 48 points (range, 24 to 48 points). The median active forward elevation was 130° (range, 65° to 170°), median external rotation was 45° (range, 5° to 80°), and median internal rotation was to T11 (range, buttock to T4). Seventy-eight shoulders demonstrated a perfect Lazarus score for radiolucency (0, indicating no radiolucency). Sixty-eight shoulders demonstrated complete osseointegration, with bone ingrowth between all of the flanges seen on coronal CT; 5 demonstrated partial osseointegration; and 10 demonstrated osteolysis around the central peg. Most radiolucent lines were in the inferior Yian zones. There were no correlations between the Yian CT scores and either the ASES or Oxford score (rho = 0.13 and 0.07, respectively).

CONCLUSIONS

Total shoulder arthroplasty with an all-polyethylene pegged glenoid component, utilizing hybrid fixation, demonstrated excellent clinical and radiographic results at the time of early follow-up. Radiolucent lines were seen most commonly around the inferior pegs of the prosthesis, and this may represent an incipient mode of failure.

LEVEL OF EVIDENCE

Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.

摘要

背景

盂肱关节假体松动仍是肩关节置换中长期失败的主要原因。本研究的目的是评估使用专为混合(生物和骨水泥)固定设计的全聚乙烯带钉盂肱关节假体后的早期临床和影像学结果。

方法

77例患者(平均年龄68.6岁)的83个肩关节接受了带钉混合固定假体(骨长入型盂肱关节和骨水泥固定钉)的全肩关节置换术。采用美国肩肘外科医师(ASES)评分和牛津肩关节评分来确定结果。对X线片和薄层计算机断层扫描(CT)图像进行详细分析,以确定凸缘间骨长入的程度以及假体-骨界面处透亮线的程度。

结果

平均随访时间为46.67个月(范围24至99个月)。在末次随访时,ASES评分中位数为97分(范围43至100分),牛津评分中位数为48分(范围24至48分)。主动前屈的中位数为130°(范围65°至170°),外旋的中位数为45°(范围5°至80°),内旋的中位数为T11(范围从臀部到T4)。78个肩关节的透亮线Lazarus评分为完美(0分,表明无透亮线)。68个肩关节显示完全骨整合,在冠状面CT上可见所有凸缘间有骨长入;5个显示部分骨整合;10个显示中央固定钉周围有骨溶解。大多数透亮线位于Yian下区。Yian CT评分与ASES评分或牛津评分之间均无相关性(相关系数分别为0.13和0.07)。

结论

采用混合固定的全聚乙烯带钉盂肱关节假体进行全肩关节置换术,在早期随访时显示出优异的临床和影像学结果。透亮线最常见于假体下固定钉周围,这可能代表一种早期失败模式。

证据水平

治疗性IV级。有关证据水平的完整描述,请参阅作者须知。