Elli Stefano, Abbruzzese Chiara, Cannizzo Luigi, Lucchini Alberto
Emergency Department and Intensive Care, San Gerardo Hospital, University of Milan-Bicocca, Monza (MB) - Italy.
J Vasc Access. 2016 Sep 21;17(5):429-34. doi: 10.5301/jva.5000583. Epub 2016 Jul 20.
To evaluate fluid reflux, when disconnecting syringe, for different needleless connectors.
Nine connectors were tested; 540 measurements were carried out.
The connectors tested showed very different performances, about reflux, on disconnection of the syringe used for flushing.The calculated reflux volumes are: Max Zero® - BD: 6.90 (±2.47) mm3; MicroClave Clear® - ICU Medical: 6.14 (±1.46) mm3; Bionecteur® - Vygon: 1.24(±0.73) mm3; Neutron® - ICU Medical: 0.12 (±0.15) mm3; SmartSite® Carefusion: 33.51 (±11.50) mm3; Safe Plus® - Cremascoli: 23.54 (±3.56) mm3; NeutraClear® - Cair: 9.36 (±1.87) mm3; NeutroX® - Cair: 0.33 (±0.31) mm3; Dasa® BTC: 2.38 (±1.67) mm3.Differences between investigated devices were statistically significant (p<0.001).
It is difficult to establish the best quality-price ratio for needleless connectors. It is important to consider several variable factors: continuous or discontinuous infusion, catheter type, usage environment and caliber of catheter used. It would therefore be useful to have an indication of the intraluminal space potentially affected by blood reflux in relation to a specific device.
Needleless connector is one of the main factors involved in keeping catheter patency. It is important to perform the best choice among the connectors available.An empirical reflux measurement, relative to the needleless connector and the catheter in use, can be obtained using an 18G cannula.
评估不同无针连接器在断开注射器时的液体回流情况。
测试了9种连接器;进行了540次测量。
所测试的连接器在用于冲洗的注射器断开连接时,在回流方面表现出非常不同的性能。计算出的回流体积分别为:Max Zero® - BD:6.90(±2.47)立方毫米;MicroClave Clear® - ICU Medical:6.14(±1.46)立方毫米;Bionecteur® - Vygon:1.24(±0.73)立方毫米;Neutron® - ICU Medical:0.12(±0.15)立方毫米;SmartSite® Carefusion:33.51(±11.50)立方毫米;Safe Plus® - Cremascoli:23.54(±3.56)立方毫米;NeutraClear® - Cair:9.36(±1.87)立方毫米;NeutroX® - Cair:0.33(±0.31)立方毫米;Dasa® BTC:2.38(±1.67)立方毫米。所研究设备之间的差异具有统计学意义(p<0.001)。
很难确定无针连接器的最佳性价比。重要的是要考虑几个可变因素:连续或间断输注、导管类型、使用环境以及所用导管的管径。因此,了解与特定设备相关的可能受血液回流影响的管腔内空间情况会很有用。
无针连接器是保持导管通畅的主要因素之一。在可用的连接器中做出最佳选择很重要。使用18G套管可以获得相对于所使用的无针连接器和导管的经验性回流测量值。