Mariyaselvam Maryanne, Hodges Emily, Richardson James, Steel Alistair, Moondi Parvez, Young Peter
The Critical Care Department, The Queen Elizabeth Hospital NHS Trust , Kings Lynn , UK.
J Med Eng Technol. 2015;39(5):259-63. doi: 10.3109/03091902.2015.1040895. Epub 2015 May 13.
Catheter-related blood stream infections (CR-BSI) account for 30% of healthcare acquired infection (HAI). Colonization of connector hubs and contaminated syringes are thought to increase the risk of CR-BSI. The Coated Antiseptic Tip (CAT) syringe was developed to decontaminate connector hubs, thereby reducing the risk of CR-BSI. Needleless valves (n = 20) and three-way connectors (n = 20) were contaminated with common critical care pathogens. At hourly intervals, CAT syringes were inserted into the connector hubs and normal saline was injected through the connector. This was repeated with control (non-coated) syringes. The internal surface of the connector hubs were swabbed at t = 0, t = 1 h and t = 4 h, inoculated onto blood agar plates and analysed by a blinded microbiologist. Growth was counted as the number of colony forming units. Baseline swabbing demonstrated 100% bacterial hub colonization in both connectors. The CAT syringe showed a significant reduction in CFU growth at 0 and 1 h compared with control syringes (p < 0.05). At 4 h, the CAT syringe completely eliminated bacterial growth in both of the connector hubs. The CAT syringe can effectively disinfect both three-way and needleless connectors.
导管相关血流感染(CR-BSI)占医疗保健相关感染(HAI)的30%。认为接头处的定植和受污染的注射器会增加CR-BSI的风险。开发了带涂层抗菌尖端(CAT)的注射器来对接头处进行消毒,从而降低CR-BSI的风险。将无针阀(n = 20)和三通接头(n = 20)用常见的重症监护病原体污染。每隔一小时,将CAT注射器插入接头处,并通过接头注入生理盐水。使用对照(无涂层)注射器重复此操作。在t = 0、t = 1小时和t = 4小时对接头处的内表面进行擦拭,接种到血琼脂平板上,并由一位不知情的微生物学家进行分析。生长情况以菌落形成单位的数量来计数。基线擦拭显示两个接头处的细菌定植率均为100%。与对照注射器相比,CAT注射器在0小时和1小时时CFU生长显著降低(p < 0.05)。在第4小时,CAT注射器完全消除了两个接头处的细菌生长。CAT注射器可有效消毒三通接头和无针接头。