Chen S N, Gauthier M, Bazalova-Carter M, Bolanos S, Glenzer S, Riquier R, Revet G, Antici P, Morabito A, Propp A, Starodubtsev M, Fuchs J
LULI-CNRS, Ecole Polytechnique, CEA: Universite Paris-Saclay, UPMC Univ Paris 06, Sorbonne Universities, F-91128 Palaiseau Cedex, France.
SLAC National Accelerator Laboratory, 2575 Sand Hill Road, Menlo Park, California 94025, USA.
Rev Sci Instrum. 2016 Jul;87(7):073301. doi: 10.1063/1.4954921.
Radiochromic films (RCF) are commonly used in dosimetry for a wide range of radiation sources (electrons, protons, and photons) for medical, industrial, and scientific applications. They are multi-layered, which includes plastic substrate layers and sensitive layers that incorporate a radiation-sensitive dye. Quantitative dose can be retrieved by digitizing the film, provided that a prior calibration exists. Here, to calibrate the newly developed EBT3 and HDv2 RCFs from Gafchromic™, we used the Stanford Medical LINAC to deposit in the films various doses of 10 MeV photons, and by scanning the films using three independent EPSON Precision 2450 scanners, three independent EPSON V750 scanners, and two independent EPSON 11000XL scanners. The films were scanned in separate RGB channels, as well as in black and white, and film orientation was varied. We found that the green channel of the RGB scan and the grayscale channel are in fact quite consistent over the different models of the scanner, although this comes at the cost of a reduction in sensitivity (by a factor ∼2.5 compared to the red channel). To allow any user to extend the absolute calibration reported here to any other scanner, we furthermore provide a calibration curve of the EPSON 2450 scanner based on absolutely calibrated, commercially available, optical density filters.
放射变色薄膜(RCF)常用于多种辐射源(电子、质子和光子)的剂量测定,应用于医学、工业和科学领域。它们是多层的,包括塑料基底层和含有辐射敏感染料的敏感层。如果存在预先校准,通过对薄膜进行数字化处理可以获取定量剂量。在此,为了校准Gafchromic™新开发的EBT3和HDv2 RCF,我们使用斯坦福医用直线加速器在薄膜中沉积不同剂量的10 MeV光子,并使用三台独立的爱普生Precision 2450扫描仪、三台独立的爱普生V750扫描仪和两台独立的爱普生11000XL扫描仪对薄膜进行扫描。薄膜在单独的RGB通道以及黑白模式下进行扫描,并且改变薄膜方向。我们发现,RGB扫描的绿色通道和灰度通道在不同型号的扫描仪上实际上相当一致,尽管这是以灵敏度降低为代价的(与红色通道相比降低了约2.5倍)。为了使任何用户能够将此处报告的绝对校准扩展到任何其他扫描仪,我们还基于绝对校准的市售光密度滤光片提供了爱普生2450扫描仪的校准曲线。