This paper presents the trouble-shooting for a very unusual stability case. Tetracaine was found unstable in neat solutions only at high concentrations, but not at low concentrations. Moreover, its stable-isotope labeled internal standard did not show similar behavior. A series of trouble-shooting experiments were conducted to uncover the root cause. Some generally applicable precautions/insights can be drawn from this investigation to avoid potential stability issues during bioanalytical method development and validation.