五种血糖检测系统在低血糖、正常血糖及高血糖范围内的综合性能评估

A Comprehensive Performance Evaluation of Five Blood Glucose Systems in the Hypo-, Eu-, and Hyperglycemic Range.

作者信息

Zijlstra Eric, Heinemann Lutz, Fischer Annelie, Kapitza Christoph

机构信息

Profil, Neuss, Germany

Profil, Neuss, Germany.

出版信息

J Diabetes Sci Technol. 2016 Nov 1;10(6):1316-1323. doi: 10.1177/1932296816668373. Print 2016 Nov.

Abstract

BACKGROUND

The objective was to evaluate the performance (in terms of accuracy, precision, and trueness) of 5 CE-certified and commercially available blood glucose (BG) systems (meters plus test strips) using an innovative clinical-experimental study design with a 3-step glucose clamp approach and frequent capillary BG measurements.

METHODS

Sixteen subjects with type 1 diabetes participated in this open label, single center trial. BG was clamped at 3 levels for 60 minutes each: 60-100-200 mg/dL. Medical staff performed regular finger pricks (up to 10 per BG level) to obtain capillary blood samples for paired BG measurements with the 5 BG systems and a laboratory method as comparison.

RESULTS

Three BG systems displayed significantly lower mean absolute relative deviations (MARD) (ACCU-Chek® Aviva Nano [5.4%], BGStar® [5.1%], iBGStar® [5.3%]) than 2 others (FreeStyle InsuLinx® [7.7%], OneTouch Verio®IQ [10.3%]). The measurement precision of all BG systems was comparable, but relative bias was also lower for the 3 systems with lower MARD (ACCU-Chek [1.3%], BGStar [-0.9%], iBGStar [1.0%]) compared with the 2 others (FreeStyle [-7.2%], OneTouch [8.9%]).

CONCLUSIONS

This 3 range glucose clamp approach enables a systematic performance evaluation of BG systems under controlled and reproducible conditions. The random error of the tested BG systems was comparable, but some showed a lower systematic error than others. These BG systems allow an accurate glucose measurement at low, normal and high BG levels.

摘要

背景

目的是采用创新的临床实验研究设计,通过三步血糖钳夹法和频繁的毛细血管血糖测量,评估5种获得CE认证的市售血糖(BG)系统(血糖仪加试纸条)的性能(准确性、精密度和真实性)。

方法

16名1型糖尿病患者参与了这项开放标签、单中心试验。血糖在3个水平上各钳夹60分钟:60 - 100 - 200mg/dL。医护人员定期进行手指采血(每个血糖水平最多10次),以获取毛细血管血样,用于与5种BG系统进行配对血糖测量,并与实验室方法进行比较。

结果

3种BG系统的平均绝对相对偏差(MARD)显著低于其他2种(拜安易®安准血糖仪[5.4%]、艾科灵睿®血糖仪[5.1%]、艾科精益®血糖仪[5.3%]),分别为(利舒坦®血糖仪[7.7%]、稳豪倍易®血糖仪[10.3%])。所有BG系统的测量精密度相当,但与其他2种系统(利舒坦[-7.2%]、稳豪[8.9%])相比,MARD较低的3种系统(拜安易[1.3%]、艾科灵睿[-0.9%]、艾科精益[1.0%])的相对偏差也较低。

结论

这种三范围血糖钳夹法能够在可控且可重复的条件下对BG系统进行系统的性能评估。被测BG系统的随机误差相当,但有些系统的系统误差低于其他系统。这些BG系统能够在低、正常和高血糖水平下进行准确的血糖测量。

相似文献

1
A Comprehensive Performance Evaluation of Five Blood Glucose Systems in the Hypo-, Eu-, and Hyperglycemic Range.
J Diabetes Sci Technol. 2016 Nov 1;10(6):1316-1323. doi: 10.1177/1932296816668373. Print 2016 Nov.
2
Clinical assessment of the accuracy of blood glucose measurement devices.
Curr Med Res Opin. 2012 Apr;28(4):525-31. doi: 10.1185/03007995.2012.673479. Epub 2012 Mar 21.
3
6
Evaluation of the FreeStyle® Libre Flash Glucose Monitoring System in Children and Adolescents with Type 1 Diabetes.
Horm Res Paediatr. 2018;89(3):189-199. doi: 10.1159/000487361. Epub 2018 Mar 27.
7
Evaluation of the effects of insufficient blood volume samples on the performance of blood glucose self-test meters.
J Diabetes Sci Technol. 2013 Nov 1;7(6):1522-9. doi: 10.1177/193229681300700612.
8
Evaluation of Accuracy of Six Blood Glucose Monitoring Systems and Modeling of Possibly Related Insulin Dosing Errors.
Diabetes Technol Ther. 2017 Oct;19(10):580-588. doi: 10.1089/dia.2016.0408. Epub 2017 Jul 13.
10
Clinical evaluation of a novel test strip technology for blood glucose monitoring: accuracy at hypoglycaemic glucose levels.
Diabetes Res Clin Pract. 2012 Dec;98(3):430-5. doi: 10.1016/j.diabres.2012.09.027. Epub 2012 Oct 9.

引用本文的文献

本文引用的文献

1
Development of the Diabetes Technology Society Blood Glucose Monitor System Surveillance Protocol.
J Diabetes Sci Technol. 2016 May 3;10(3):697-707. doi: 10.1177/1932296815614587. Print 2016 May.
3
How to Assess the Quality of Glucose Clamps? Evaluation of Clamps Performed With ClampArt, a Novel Automated Clamp Device.
J Diabetes Sci Technol. 2015 Jul;9(4):792-800. doi: 10.1177/1932296815576957. Epub 2015 Apr 7.
9
Clinical assessment of the accuracy of blood glucose measurement devices.
Curr Med Res Opin. 2012 Apr;28(4):525-31. doi: 10.1185/03007995.2012.673479. Epub 2012 Mar 21.
10
Acute hypoglycemia decreases myocardial blood flow reserve in patients with type 1 diabetes mellitus and in healthy humans.
Circulation. 2011 Oct 4;124(14):1548-56. doi: 10.1161/CIRCULATIONAHA.110.992297. Epub 2011 Sep 12.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验