Determination of novel brominated flame retardants and polybrominated diphenyl ethers in serum using gas chromatography-mass spectrometry with two simplified sample preparation procedures.

作者信息

Gao Le, Li Jian, Wu Yandan, Yu Miaohao, Chen Tian, Shi Zhixiong, Zhou Xianqing, Sun Zhiwei

机构信息

School of Public Health, Capital Medical University, Beijing, 100069, China.

Beijing Key Laboratory of Environmental Toxicology, Capital Medical University, Beijing, 100069, China.

出版信息

Anal Bioanal Chem. 2016 Nov;408(27):7835-7844. doi: 10.1007/s00216-016-9887-z. Epub 2016 Sep 10.

Abstract

Two simple and efficient pretreatment procedures have been developed for the simultaneous extraction and cleanup of six novel brominated flame retardants (NBFRs) and eight common polybrominated diphenyl ethers (PBDEs) in human serum. The first sample pretreatment procedure was a quick, easy, cheap, effective, rugged, and safe (QuEChERS)-based approach. An acetone/hexane mixture was employed to isolate the lipid and analytes from the serum with a combination of MgSO and NaCl, followed by a dispersive solid-phase extraction (d-SPE) step using C18 particles as a sorbent. The second sample pretreatment procedure was based on solid-phase extraction. The sample extraction and cleanup were conducted directly on an Oasis HLB SPE column using 5 % aqueous isopropanol, concentrated sulfuric acid, and 10 % aqueous methanol, followed by elution with dichloromethane. The NBFRs and PBDEs were then detected using gas chromatography-negative chemical ionization mass spectrometry (GC-NCI MS). The methods were assessed for repeatability, accuracy, selectivity, limits of detection (LODs), and linearity. The results of spike recovery experiments in fetal bovine serum showed that average recoveries ranged from 77.9 % to 128.8 % with relative standard deviations (RSDs) from 0.73 % to 12.37 % for most of the analytes. The LODs for the analytes in fetal bovine serum ranged from 0.3 to 50.8 pg/mL except for decabromodiphenyl ethane. The proposed method was successfully applied to the determination of the 14 brominated flame retardants in human serum. The two pretreatment procedures described here are simple, accurate, and precise, and are suitable for the routine analysis of human serum. Graphical Abstract Workflow of a QuEChERS-based approach (top) and an SPE-based approach (bottom) for the detection of PBDEs and NBFRs in serum.

摘要

已开发出两种简单有效的预处理程序,用于同时提取和净化人血清中的六种新型溴化阻燃剂(NBFRs)和八种常见的多溴二苯醚(PBDEs)。第一种样品预处理程序是基于快速、简便、廉价、有效、耐用且安全(QuEChERS)的方法。使用丙酮/己烷混合物,结合硫酸镁和氯化钠从血清中分离脂质和分析物,随后以C18颗粒作为吸附剂进行分散固相萃取(d-SPE)步骤。第二种样品预处理程序基于固相萃取。样品的萃取和净化直接在Oasis HLB固相萃取柱上进行,使用5%的异丙醇水溶液、浓硫酸和10%的甲醇水溶液,随后用二氯甲烷洗脱。然后使用气相色谱-负化学电离质谱(GC-NCI MS)检测NBFRs和PBDEs。对这些方法的重复性、准确性、选择性、检测限(LODs)和线性进行了评估。胎牛血清中加标回收实验的结果表明,大多数分析物的平均回收率在77.9%至128.8%之间,相对标准偏差(RSDs)在0.73%至12.37%之间。除十溴二苯乙烷外,胎牛血清中分析物的检测限在0.3至50.8 pg/mL之间。所提出的方法成功应用于人血清中14种溴化阻燃剂的测定。这里描述的两种预处理程序简单、准确且精确,适用于人血清的常规分析。图形摘要基于QuEChERS的方法(顶部)和基于固相萃取的方法(底部)检测血清中PBDEs和NBFRs的工作流程。