双能X线吸收法测定骨矿物质的质量保证与精密度

Quality assurance and precision of dual photon absorptiometry in bone mineral measurement.

作者信息

Tzen K Y, Wu C

出版信息

Zhonghua Yi Xue Za Zhi (Taipei). 1989 Jan;43(1):1-8.

PMID:2766058
Abstract

Dual photon absorptiometry (DPA) has evolved as a powerful technique in measuring bone mineral density (BMD) to assess osteoporosis and metabolic bone diseases in recent years. We introduced the first installation of DPA instrument into this country in Aug. 1987. Since the comparison among longitudinal BMD values influences the diagnosis, monitoring and medical decision-making, high precision of the system performance and strict quality assurance program are prerequisite for judgment of the difference between two successive measurements. However, the automation and black-box software design of the instrument make users gain very little knowledge how the instruments acquire the long-term system reliability. The first part of this study was designed to establish a standardized procedure to evaluate the long-term quality of the DPA instrument. Our instrument was proved to have stable sensitivity, resolution, accuracy and reproducibility in the four months of study. Three subjects with different BMD levels were scanned to evaluate the intraobserver and interobserver errors in BMD analysis expected in our laboratory (2.2 and 1.6% at 95% confidence level, respectively). The results indicated that reanalyzing the same scan by either of the two observers would produce a BMD precision error within 2.2% at a 95% confidence level. Serial scans of 7 normal volunteers within five weeks were then conducted to evaluate the overall precision of the BMD measurements. Our findings indicated that two scans would be considered as having a real BMD change if they differ by more than 3.3% at a 95% confidence level. There are at least five sets of DPA instruments that have been installed in Taiwan presently.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

近年来,双能X线吸收法(DPA)已发展成为一种用于测量骨矿物质密度(BMD)以评估骨质疏松症和代谢性骨病的强大技术。1987年8月,我们将第一台DPA仪器引入该国。由于纵向BMD值之间的比较会影响诊断、监测和医疗决策,因此系统性能的高精度和严格的质量保证程序是判断两次连续测量之间差异的先决条件。然而,仪器的自动化和黑箱软件设计使用户对仪器如何获得长期系统可靠性了解甚少。本研究的第一部分旨在建立一个标准化程序来评估DPA仪器的长期质量。在四个月的研究中,我们的仪器被证明具有稳定的灵敏度、分辨率、准确性和可重复性。扫描了三名BMD水平不同的受试者,以评估我们实验室预期的BMD分析中的观察者内和观察者间误差(在95%置信水平下分别为2.2%和1.6%)。结果表明,两名观察者中的任何一名重新分析相同的扫描,在95%置信水平下,BMD精度误差将在2.2%以内。然后对7名正常志愿者在五周内进行连续扫描,以评估BMD测量的总体精度。我们的研究结果表明,如果两次扫描在95%置信水平下相差超过3.3%,则可认为存在真正的BMD变化。目前台湾至少已安装了五套DPA仪器。(摘要截断于250字)

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验