脉冲重复频率对三维能量多普勒定量分析的影响

Influence of Pulse Repetition Frequency on 3-D Power Doppler Quantification.

作者信息

Soares Carlos A M, Pavan Theo Z, Miyague Andre H, Kudla Marek, Martins Wellington P

机构信息

Department of Obstetrics and Gynecology, Ribeirao Preto Medical School, University of Sao Paulo (DGO-FMRP-USP), Ribeirao Preto, Brazil.

Department of Physics, School of Philosophy, Sciences and Letters of Ribeirao Preto, University of Sao Paulo (FFCLRP-USP), Ribeirao Preto, Brazil.

出版信息

Ultrasound Med Biol. 2016 Dec;42(12):2887-2892. doi: 10.1016/j.ultrasmedbio.2016.08.015. Epub 2016 Sep 24.

Abstract

Three-dimensional power Doppler quantification has limited application because of its high dependency on attenuation. The purpose of the study described here was to assess if different degrees of attenuation, depending on pulse repetition frequency (PRF) adjustment, alter 3-D power Doppler quantification in a region of 100% moving blood when using vascularization index, flow index and vascularization flow index (VFI). A cubic-shaped gelatin phantom with a 1.8-mm-internal-diameter silicon tube was used. The tube, placed at 45° to the phantom's surface, was filled with blood-mimicking fluid with as constant maximum velocity of 30 cm/s. Two different attenuation blocks (low and high attenuation) were alternatively placed between the phantom and the transvaginal transducer. One single observer acquired 10 data sets for each PRF level from 0.3 to 7.5 kHz, using the high- and low-attenuation blocks, for a total of 200 3-D power Doppler data sets. We assessed VFI from 1.5-mm-diameter spherical samples, virtually placed inside the tube, always at the same position. No difference was noted between high- and low-attenuation VFI values when using a PRF of 0.3 kHz. As PRF increased, it was observed that VFI quantification progressively differed between low and high attenuation. Also, a slope on VFI values for both high- and low-attenuation models could be observed when increasing PRF, particularly above 4.0 kHz. We concluded that PRF adjustment is very relevant when using VFI to quantify 3-D power Doppler signal.

摘要

三维能量多普勒定量分析的应用有限,因为它高度依赖于衰减。本文所述研究的目的是评估在使用血管化指数、血流指数和血管化血流指数(VFI)时,根据脉冲重复频率(PRF)调整而产生的不同程度的衰减,是否会改变100%流动血液区域的三维能量多普勒定量分析。使用了一个内部直径为1.8毫米的硅管的立方体明胶模型。该管与模型表面呈45°放置,充满了模拟血液的流体,其恒定最大速度为30厘米/秒。在模型和经阴道换能器之间交替放置两个不同的衰减块(低衰减和高衰减)。一名观察者针对从0.3至7.5千赫的每个PRF水平,使用高衰减和低衰减块获取了10组数据集,总共200个三维能量多普勒数据集。我们从虚拟放置在管内、始终处于同一位置的直径为1.5毫米的球形样本中评估VFI。当使用0.3千赫的PRF时,高衰减和低衰减的VFI值之间未观察到差异。随着PRF的增加,可以观察到低衰减和高衰减之间的VFI定量逐渐不同。此外,当增加PRF时,尤其是在4.0千赫以上时,可以观察到高衰减和低衰减模型的VFI值都有一个斜率。我们得出结论,在使用VFI对三维能量多普勒信号进行定量分析时,PRF调整非常重要。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验