Initial clinical results with the ThermoCool® SmartTouch® Surround Flow catheter.

机构信息

Cardiology Clinical Academic Group, St. George's University Hospitals NHS Foundation Trust, St. George's, University of London, London, UK.

National Heart & Lung Institute, Imperial College London, London, UK.

出版信息

Europace. 2017 Aug 1;19(8):1317-1321. doi: 10.1093/europace/euw177.

Abstract

AIM

The Biosense Webster ThermoCool® SmartTouch® Surround Flow (STSF) catheter is a recently developed ablation catheter incorporating Surround Flow (SF) technology to ensure efficient cooling and force sensing to quantify tissue contact. In our unit, it superseded the ThermoCool® SF catheter from the time of its introduction in May 2015.

METHODS AND RESULTS

Procedure-related data were collected prospectively for the first 100 ablation procedures performed in our department using the STSF catheter. From a database of 654 procedures performed in our unit using the SF catheter, we selected one to match each STSF procedure, matching for procedure type, operator experience, patient age, and gender. The groups were well matched for patient age, gender, and procedure type. Procedure duration was similar in both groups (mean 225.5 vs. 221.4 min, IQR 106.5 vs. 91.5, P = 0.55), but fluoroscopy duration was shorter in the STSF group (mean 25.8 vs. 30.0, IQR 19.6 vs. 18.5, P = 0.03). No complication occurred in the STSF group. Complications occurred in two cases in the SF group (one pericardial effusion requiring drainage and one need for permanent pacing). Complete procedural success was achieved in 98 cases in the STSF group and 94 cases in the SF group (P = 0.15). The composite endpoint of procedure failure or acute complication was less common in the STSF group (2 vs. 8, P = 0.05).

CONCLUSION

The STSF catheter is safe and effective in treating a range of arrhythmias. Compared with the SF catheter, it shows a trend towards improved safety-efficacy balance.

摘要

目的

Biosense Webster ThermoCool® SmartTouch® Surround Flow(STSF)导管是一种最近开发的消融导管,结合了 Surround Flow(SF)技术,以确保有效的冷却和力感应,以量化组织接触。在我们的单位,它从 2015 年 5 月推出以来取代了 ThermoCool® SF 导管。

方法和结果

前瞻性地收集了我们部门使用 STSF 导管进行的前 100 次消融程序的程序相关数据。在我们的单位使用 SF 导管进行的 654 次程序中,我们选择了一个与每个 STSF 程序相匹配的程序,匹配程序类型、操作人员经验、患者年龄和性别。两组患者的年龄、性别和程序类型匹配良好。两组的手术时间相似(平均值分别为 225.5 分钟和 221.4 分钟,IQR 分别为 106.5 分钟和 91.5 分钟,P = 0.55),但 STSF 组的透视时间更短(平均值分别为 25.8 分钟和 30.0 分钟,IQR 分别为 19.6 分钟和 18.5 分钟,P = 0.03)。STSF 组无并发症发生。SF 组有两例并发症(一例心包积液需要引流,一例需要永久性起搏)。STSF 组有 98 例程序完全成功,SF 组有 94 例(P = 0.15)。STSF 组的程序失败或急性并发症复合终点较少(2 例 vs. 8 例,P = 0.05)。

结论

STSF 导管在治疗多种心律失常方面是安全有效的。与 SF 导管相比,它显示出改善安全性-疗效平衡的趋势。