We established a mathematical method to precisely calculate the radiographic anteversion (RA) and radiographic inclination (RI) angles of the acetabular cup based on anterior-posterior (AP) pelvic radiographs after total hip arthroplasty. Using Mathematica software, a mathematical model for an oblique cone was established to simulate how AP pelvic radiographs are obtained and to address the relationship between the two-dimensional and three-dimensional geometry of the opening circle of the cup. In this model, the vertex was the X-ray beam source, and the generatrix was the ellipse in radiographs projected from the opening circle of the acetabular cup. Using this model, we established a series of mathematical formulas to reveal the differences between the true RA and RI cup angles and the measurements results achieved using traditional methods and AP pelvic radiographs and to precisely calculate the RA and RI cup angles based on post-operative AP pelvic radiographs. Statistical analysis indicated that traditional methods should be used with caution if traditional measurements methods are used to calculate the RA and RI cup angles with AP pelvic radiograph. The entire calculation process could be performed by an orthopedic surgeon with mathematical knowledge of basic matrix and vector equations.
我们建立了一种数学方法,用于在全髋关节置换术后基于前后位(AP)骨盆X线片精确计算髋臼杯的放射学前倾角(RA)和放射学倾斜角(RI)。使用Mathematica软件,建立了一个斜圆锥的数学模型,以模拟如何获得AP骨盆X线片,并处理髋臼杯开口圆的二维和三维几何形状之间的关系。在该模型中,顶点为X射线束源,母线为从髋臼杯开口圆投影到X线片中的椭圆。利用该模型,我们建立了一系列数学公式,以揭示髋臼杯的真实RA和RI角度与使用传统方法和AP骨盆X线片获得的测量结果之间的差异,并基于术后AP骨盆X线片精确计算髋臼杯的RA和RI角度。统计分析表明,如果使用传统测量方法通过AP骨盆X线片计算髋臼杯的RA和RI角度,应谨慎使用传统方法。整个计算过程可由具有基本矩阵和向量方程数学知识的骨科医生完成。