Sim L
Radiation Oncology Mater Centre, Princess Alexandra Hospital, South Brisbane, QLD, 4101, Australia.
Australas Phys Eng Sci Med. 2016 Dec;39(4):863-870. doi: 10.1007/s13246-016-0480-8. Epub 2016 Oct 11.
The purpose of this study was to evaluate if MED610 3D printed material can be used as a surrogate for acrylic in the manufacturing of a replacement insert used in an eye plaque brachytherapy applicator. Measurement of the dose distributions from a standard acrylic insert were compared with dose obtained from MED610 3D printed replica using GafChromic EBT3 films. The study used a 15 mm Radiation Oncology Physics and Engineering Services, Australia (ROPES) type eye plaque applicator loaded with I-125 (model 6711) seeds. GafChromic EBT3 films were placed in a solid water phantom and dose distributions were measured three-dimensionally both along and perpendicular to a loaded ROPES eye plaque's central axis (CAX). Each measurement was performed with the stainless steel plaque backing attached to the eye plaque, to assess the variability of the dose distributions between the acrylic and MED 610 insert. Results of dose along the central axis were compared between acrylic and MED610 insert and the results found agreement within 1.5 %. Off-axis profiles were also compared between the acrylic insert and MED610 and were found to agree to within 7 % in the central 15 mm width centred on CAX at depths ranging from z = 2 mm to z = 8 mm in 2 mm increments. The aim of this investigation was to verify the consistency between doses profiles over a range of clinically relevant depths for a 15 mm loaded ROPES plaque using acrylic versus MED610 material. The results show an agreement between experimental measurements given the film uncertainty of 7 %.
本研究的目的是评估MED610 3D打印材料是否可在制造眼贴近距离放射治疗施源器中使用的替换插入物时替代丙烯酸材料。使用GafChromic EBT3薄膜,将标准丙烯酸插入物的剂量分布测量结果与MED610 3D打印复制品获得的剂量进行比较。该研究使用了一个装有I-125(型号6711)种子的15毫米澳大利亚放射肿瘤学物理与工程服务(ROPES)型眼贴施源器。将GafChromic EBT3薄膜放置在固体水模体中,沿加载的ROPES眼贴的中心轴(CAX)及其垂直方向进行三维剂量分布测量。每次测量时,不锈钢贴板背衬都连接到眼贴上,以评估丙烯酸和MED 610插入物之间剂量分布的变异性。比较了丙烯酸和MED610插入物沿中心轴的剂量结果,发现结果在1.5%以内相符。还比较了丙烯酸插入物和MED610的离轴分布,发现在以CAX为中心、宽度为15毫米的中心区域,在深度从z = 2毫米到z = 8毫米、增量为2毫米的范围内,两者相符程度在7%以内。本研究的目的是验证对于一个装有15毫米种子的ROPES贴板,在一系列临床相关深度下,使用丙烯酸材料与MED610材料时剂量分布的一致性。结果表明,考虑到薄膜7%的不确定性,实验测量结果相符。