Accu-Chek inform II即时血糖仪的大规模性能评估。

Large-scale performance evaluation of Accu-Chek inform II point-of-care glucose meters.

作者信息

Jeong Tae-Dong, Cho Eun-Jung, Ko Dae-Hyun, Lee Woochang, Chun Sail, Hong Ki-Sook, Min Won-Ki

机构信息

a Department of Laboratory Medicine , Ewha Womans University School of Medicine , Seoul , Korea.

b Department of Laboratory Medicine , University of Ulsan College of Medicine and Asan Medical Center , Seoul , Korea.

出版信息

Scand J Clin Lab Invest. 2016 Dec;76(8):657-663. doi: 10.1080/00365513.2016.1233575. Epub 2016 Oct 14.

Abstract

BACKGROUND

The aim of this study was to report the experience of large-scale performance evaluation of 238 Accu-Chek Inform II point-of-care (POC) glucose meters in a single medical setting.

METHODS

The repeatability of 238 POC devices, the within-site imprecision of 12 devices, and the linearity of 49 devices were evaluated using glucose control solutions. The glucose results of 24 POC devices and central laboratory were compared using patient samples.

RESULTS

Mean concentration of control solutions was 2.39 mmol/L for Level 1 and 16.52 mmol/L for Level 2. The pooled repeatability coefficient of variation (CV) of the 238 devices was 2.0% for Level 1 and 1.6% for Level 2. The pooled within-site imprecision CV and reproducibility CV of the 12 devices were 2.7% and 2.7% for Level 1, and 1.9%, and 1.9% for Level 2, respectively. The test results of all 49 devices were linear within analytical measurement range from 1.55-31.02 mmol/L. The correlation coefficient for individual POC devices ranged from 0.9967-0.9985. The total correlation coefficient for the 24 devices was 0.998.

CONCLUSIONS

The Accu-Chek Inform II POC blood glucose meters performed well in terms of precision, linearity, and correlation evaluations. Consensus guidelines for the large-scale performance evaluations of POC devices are required.

摘要

背景

本研究旨在报告在单一医疗环境中对238台罗氏卓越型第二代即时检测(POC)血糖仪进行大规模性能评估的经验。

方法

使用葡萄糖控制溶液评估238台POC设备的重复性、12台设备的现场不精密度以及49台设备的线性。使用患者样本比较24台POC设备与中心实验室的血糖检测结果。

结果

1级控制溶液的平均浓度为2.39 mmol/L,2级为16.52 mmol/L。238台设备的合并重复性变异系数(CV)在1级时为2.0%,在2级时为1.6%。12台设备的合并现场不精密度CV和再现性CV在1级时分别为2.7%和2.7%,在2级时分别为1.9%和1.9%。所有49台设备的检测结果在1.55 - 31.02 mmol/L的分析测量范围内呈线性。各POC设备的相关系数在0.9967 - 0.9985之间。24台设备的总相关系数为0.998。

结论

罗氏卓越型第二代即时检测血糖仪在精密度、线性和相关性评估方面表现良好。需要制定POC设备大规模性能评估的共识指南。