Optical non-contact pH measurement in cell culture with sterilizable, modular parts.

作者信息

Kattipparambil Rajan Dhanesh, Patrikoski Mimmi, Verho Jarmo, Sivula Jyrki, Ihalainen Heimo, Miettinen Susanna, Lekkala Jukka

机构信息

Department of Automation Science and Engineering, Tampere University of Technology, Tampere, Finland.

Adult Stem Cell Group, BioMediTech, Institute of Biosciences and Medical Technology, University of Tampere, Tampere, Finland; Science Center, Tampere University Hospital, Tampere, Finland.

出版信息

Talanta. 2016 Dec 1;161:755-761. doi: 10.1016/j.talanta.2016.09.021. Epub 2016 Sep 8.

Abstract

A non-contact real time pH measurement using fully modular optical parts is described for phenol-red medium cell cultures. The modular parts can be sterilized, and once the measurement is started at the beginning of culture, no recalibration or maintenance is needed till the end of the culture. Measurements can be carried out without any special manual attention. The modular assembly of LED and sensor cassettes is unique, robust, reusable and reproducible. pH is measured in an intact closed flow system, without wasting any culture medium. A special pump encapsulation enables the system to be effortlessly functional in extremely humid incubator environments. This avoids lengthy sample tubings in and out of the incubator, associated large temperature changes and CO buffering issues. A new correction model to compensate errors caused e.g. by biolayers in spectrometric pH measurement is put-forward, which improves the accuracy of pH estimation significantly. The method provides resolution down to 0.1 pH unit in physiological pH range with mean absolute error 0.02.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验