Kwon Young-Min, Tsai Tsung-Yuan, Leone William A, Liow Ming Han Lincoln
Department of Orthopaedic Surgery, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts.
J Arthroplasty. 2017 Mar;32(3):996-1000. doi: 10.1016/j.arth.2016.08.040. Epub 2016 Sep 28.
Currently, no serum metal ion threshold exists to identify adverse tissue reactions in total hip arthroplasty (THA) patients with taper corrosion. Our study aims to investigate the sensitivity and specificity of serum metal ions in detecting taper corrosion related pseudotumors in patients with dual taper modular THA.
A total of 148 patients with dual taper modular THA were investigated: (1) 90 patients with pseudotumors detected with metal artifact reduction sequence-magnetic resonance imaging (MARS-MRI) and (2) 58 patients without pseudotumors on MARS-MRI. Receiver operating characteristic curves were constructed to determine the sensitivity and specificity using different metal ion thresholds. The severity of intraoperative tissue damage was correlated with preoperative metal ion levels.
Pseudotumor was associated with higher cobalt (5.0 μg/L vs 3.7 μg/L, P < .01) and Co/Cr ratio (6.0 vs 3.7, P < .01). The sensitivity and specificity for cobalt level of 2.8 μg/L and Co/Cr ratio of 3.8 in detecting taper corrosion-related pseudotumors on MARS-MRI was 88% and 32% and 70% and 50%, respectively. Higher intraoperative tissue damage grades demonstrated significantly higher Co/Cr ratios (8.6 vs 3.4, P = .03).
Although metal ion levels alone should not be relied on as the sole parameter to determine revision surgery, cobalt level >2.8 μg/L and the Co/Cr ratio >3.8 are useful clinical diagnostic adjuncts in the systematic clinical evaluation for taper corrosion-related adverse tissue reactions in patients with dual modular taper THA.
目前,尚无血清金属离子阈值可用于识别全髋关节置换术(THA)中出现锥形腐蚀的患者的不良组织反应。我们的研究旨在探讨血清金属离子在检测双锥形模块化THA患者中与锥形腐蚀相关假肿瘤方面的敏感性和特异性。
共对148例双锥形模块化THA患者进行了研究:(1)90例通过金属伪影减少序列磁共振成像(MARS-MRI)检测出有假肿瘤的患者;(2)58例MARS-MRI检查未发现假肿瘤的患者。构建受试者工作特征曲线,以确定使用不同金属离子阈值时的敏感性和特异性。术中组织损伤的严重程度与术前金属离子水平相关。
假肿瘤与较高的钴含量(5.0μg/L对3.7μg/L,P<.01)和钴/铬比率(6.0对3.7,P<.01)相关。MARS-MRI上检测锥形腐蚀相关假肿瘤时,钴水平为2.8μg/L和钴/铬比率为3.8时的敏感性和特异性分别为88%和32%以及70%和50%。术中组织损伤等级越高,钴/铬比率越高(8.6对3.4,P=.03)。
虽然不应仅依靠金属离子水平作为决定翻修手术的唯一参数,但钴水平>2.8μg/L和钴/铬比率>3.8在对双模块化锥形THA患者中与锥形腐蚀相关的不良组织反应进行系统临床评估时,是有用的临床诊断辅助指标。