金属离子水平对原发性金属对聚乙烯全髋关节置换术中头颈部锥形腐蚀所致不良局部组织反应的预测的敏感性和特异性。

Sensitivity and Specificity of Metal Ion Level in Predicting Adverse Local Tissue Reactions Due to Head-Neck Taper Corrosion in Primary Metal-on-Polyethylene Total Hip Arthroplasty.

机构信息

Department of Orthopaedic Surgery, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts.

出版信息

J Arthroplasty. 2018 Sep;33(9):3025-3029. doi: 10.1016/j.arth.2018.04.007. Epub 2018 Apr 16.

Abstract

BACKGROUND

There is increasing concern regarding the occurrence of adverse local tissue reactions (ALTRs) in metal-on-polyethylene (MoP) total hip arthroplasty (THA) patients due to mechanically assisted crevice corrosion at the head-neck modular taper junction. The utility of metal ion levels in evaluating MoP patients with head-neck taper has not been fully characterized. This study aimed to investigate the sensitivity and specificity of serum metal ions as a predictor of ALTR in MoP THA patients.

METHODS

A total of 62 MoP THA patients in 2 cohorts were investigated: (1) ALTR on metal artifact reduction sequence magnetic resonance imaging (n = 43) and (2) without ALTR on metal artifact reduction sequence magnetic resonance imaging (n = 19). Receiver-operating characteristic curve was constructed to determine the sensitivity and specificity using different metal ion thresholds and the optimal metal ion cutoff values.

RESULTS

The presence of ALTR was associated with significantly higher cobalt (8.9 ng/mL vs 0.3 ng/mL, P < .001), chromium (2.26 ng/mL vs 0.21 ng/mL, P < .001), and Co/Cr ratio (5.9 vs 2.0, P < .001). The sensitivity and specificity for cobalt (1 ng/mL) was 95% and 94% and for Co/Cr ratio (2) was 83% and 72%, respectively.

CONCLUSIONS

Although metal ion levels alone should not be relied on as the sole parameter to determine revision surgery, serum cobalt level of >1 ng/mL (1 ppb) and Co/Cr ratio >2 thresholds provide evidence-based practical information for surgeons when interpreting metal ion levels in MoP THA patients for clinically relevant head-neck taper corrosion.

摘要

背景

由于在头颈连接的机械辅助缝隙腐蚀,金属对聚乙烯(MoP)全髋关节置换(THA)患者中越来越多地出现不良局部组织反应(ALTR),这引起了人们的关注。金属离子水平在评估 MoP 患者头颈锥度方面的效用尚未得到充分描述。本研究旨在研究血清金属离子作为 MoP THA 患者 ALTR 预测因子的敏感性和特异性。

方法

对两个队列中的 62 例 MoP THA 患者进行了研究:(1)磁共振成像金属伪影减少序列(metal artifact reduction sequence magnetic resonance imaging,MARS-MRI)上的 ALTR(n=43);(2)MARS-MRI 上无 ALTR(n=19)。构建了受试者工作特征曲线,以确定不同金属离子阈值和最佳金属离子截断值的敏感性和特异性。

结果

ALTR 的存在与钴(8.9ng/mL 比 0.3ng/mL,P<.001)、铬(2.26ng/mL 比 0.21ng/mL,P<.001)和 Co/Cr 比值(5.9 比 2.0,P<.001)显著升高相关。钴(1ng/mL)的敏感性和特异性分别为 95%和 94%,Co/Cr 比值(2)的敏感性和特异性分别为 83%和 72%。

结论

尽管金属离子水平本身不应该作为唯一参数来确定翻修手术,但血清钴水平>1ng/mL(1ppb)和 Co/Cr 比值>2 阈值为外科医生在解释 MoP THA 患者的金属离子水平时提供了循证实用信息,以评估与临床相关的头颈锥度腐蚀。