Oechsner Markus, Chizzali Barbara, Devecka Michal, Combs Stephanie Elisabeth, Wilkens Jan Jakob, Duma Marciana Nona
Department of Radiation Oncology, Klinikum rechts der Isar, Technical University of Munich, Ismaninger Str. 22, 81675, Munich, Germany.
Institute of Innovative Radiotherapy (iRT), Helmholtz Zentrum München, Ingolstädter Landstraße 1, Munich, Germany.
Radiat Oncol. 2016 Oct 26;11(1):142. doi: 10.1186/s13014-016-0720-9.
The aim of this study was to analyze differences in couch shifts (setup errors) resulting from image registration of different CT datasets with free breathing cone beam CTs (FB-CBCT). As well automatic as manual image registrations were performed and registration results were correlated to tumor characteristics.
FB-CBCT image registration was performed for 49 patients with lung lesions using slow planning CT (PCT), average intensity projection (AIP), maximum intensity projection (MIP) and mid-ventilation CTs (MidV) as reference images. Both, automatic and manual image registrations were applied. Shift differences were evaluated between the registered CT datasets for automatic and manual registration, respectively. Furthermore, differences between automatic and manual registration were analyzed for the same CT datasets. The registration results were statistically analyzed and correlated to tumor characteristics (3D tumor motion, tumor volume, superior-inferior (SI) distance, tumor environment).
Median 3D shift differences over all patients were between 0.5 mm (AIPvsMIP) and 1.9 mm (MIPvsPCT and MidVvsPCT) for the automatic registration and between 1.8 mm (AIPvsPCT) and 2.8 mm (MIPvsPCT and MidVvsPCT) for the manual registration. For some patients, large shift differences (>5.0 mm) were found (maximum 10.5 mm, automatic registration). Comparing automatic vs manual registrations for the same reference CTs, ∆AIP achieved the smallest (1.1 mm) and ∆MIP the largest (1.9 mm) median 3D shift differences. The standard deviation (variability) for the 3D shift differences was also the smallest for ∆AIP (1.1 mm). Significant correlations (p < 0.01) between 3D shift difference and 3D tumor motion (AIPvsMIP, MIPvsMidV) and SI distance (AIPvsMIP) (automatic) and also for 3D tumor motion (∆PCT, ∆MidV; automatic vs manual) were found.
Using different CT datasets for image registration with FB-CBCTs can result in different 3D couch shifts. Manual registrations achieved partly different 3D shifts than automatic registrations. AIP CTs yielded the smallest shift differences and might be the most appropriate CT dataset for registration with 3D FB-CBCTs.
本研究旨在分析不同CT数据集与自由呼吸锥形束CT(FB-CBCT)进行图像配准所导致的治疗床移位(摆位误差)差异。同时进行了自动和手动图像配准,并将配准结果与肿瘤特征相关联。
对49例肺部病变患者,使用慢扫描计划CT(PCT)、平均强度投影(AIP)、最大强度投影(MIP)和呼气中期CT(MidV)作为参考图像进行FB-CBCT图像配准。自动和手动图像配准均被应用。分别评估自动和手动配准的已配准CT数据集之间的移位差异。此外,还分析了相同CT数据集自动和手动配准之间的差异。对配准结果进行统计学分析,并与肿瘤特征(三维肿瘤运动、肿瘤体积、上下(SI)距离、肿瘤周围环境)相关联。
所有患者自动配准的三维移位差异中位数在0.5毫米(AIP与MIP)至1.9毫米(MIP与PCT以及MidV与PCT)之间,手动配准的在1.8毫米(AIP与PCT)至2.8毫米(MIP与PCT以及MidV与PCT)之间。对于部分患者,发现了较大的移位差异(>5.0毫米)(自动配准最大为10.5毫米)。比较相同参考CT的自动与手动配准,∆AIP的三维移位差异中位数最小(1.1毫米),∆MIP最大(1.9毫米)。三维移位差异的标准差(变异性)对于∆AIP也最小(1.1毫米)。发现三维移位差异与三维肿瘤运动(AIP与MIP、MIP与MidV)以及SI距离(AIP与MIP)(自动)之间存在显著相关性(p < 0.01),对于三维肿瘤运动(∆PCT、∆MidV;自动与手动)也存在显著相关性。
使用不同的CT数据集与FB-CBCT进行图像配准会导致不同的三维治疗床移位。手动配准与自动配准在部分三维移位上有所不同。AIP CT产生的移位差异最小,可能是与三维FB-CBCT配准最合适的CT数据集。