[鼻振荡阻力的测量]

[Measurement of nasal oscillatory resistance].

作者信息

Petro W, Birkenbihl P

出版信息

Pneumologie. 1989 Jul;43(7):387-91.

PMID:2780533
Abstract

The usefulness of a new forced oscillation measuring technique for establishing the airway resistance measured nasally, was investigated in 40 patients. Forced oscillation measurements of airway resistance employing the custo vit and Siregnost FD 5, the interruptor technique and classical rhinomanometry for the measurement of nasal resistance, were compared. The correlation of the data of the nasal and bronchial resistance is significant between custo vit on the one hand, and the Siregnost FD 5 and rhinometry on the other. The forced oscillation technique employing the custo vit and the use of a facemask under conditions of normal breathing is, accordingly, a reliable method for determining nasal resistance. The spectrum of application of the equipment is thus expanded from spirometry to rhinomanometry, and its usefulness in nasal provocation testing confirmed.

摘要

在40名患者中研究了一种用于测量经鼻气道阻力的新型强迫振荡测量技术的实用性。比较了使用custo vit和Siregnost FD 5进行气道阻力的强迫振荡测量、用于测量鼻阻力的阻断器技术和经典鼻阻力测量法。一方面,custo vit与另一方面的Siregnost FD 5和鼻阻力测量法之间,鼻阻力和支气管阻力数据的相关性显著。因此,在正常呼吸条件下使用custo vit的强迫振荡技术以及使用面罩是确定鼻阻力的可靠方法。该设备的应用范围因此从肺量测定扩展到鼻阻力测量,并证实了其在鼻激发试验中的实用性。