Rizvi Bisma, Da Silva Eric, Slatkovska Luba, Cheung Angela M, Tavakkoli Jahan, Pejović-Milić Ana
Ryerson University, Department of Physics, Toronto, Ontario M5B 2K3, Canada.
University Health Network, Osteoporosis Program, Toronto, Ontario M5G 2C4, Canada and University of Toronto, Centre of Excellence in Skeletal Health Assessment, Toronto, Ontario M5G 2C4, Canada.
Med Phys. 2016 Nov;43(11):5817. doi: 10.1118/1.4963805.
Bone quantity, as determined by the current gold standard, dual energy X-ray absorptiometry (DXA), through measured areal bone mineral density (aBMD), is subject to positive biases if bone strontium levels are high. This is of particular concern for populations administered strontium-based compounds for the treatment of osteoporosis. This study investigated the dependence of bone mineral density (BMD) determinations, and associated ultrasound-determined indices, obtained by quantitative ultrasound (QUS), on bone strontium content using a new generation of trabecular bone-mimicking phantoms.
A new generation of bone-mimicking phantoms, consisting of hydroxyapatite (HA) and gelatin, was developed. Castor oil layers were included in these phantoms to create a multilayer bone-mimicking phantom. These phantoms were prepared using a bone mineral fraction consisting of varying strontium concentrations in the range of 0-2.5% mol/mol as strontium-substituted HA. The effect of varying bone strontium content on determined quality indices was evaluated based on determined speed of sound (SOS), broadband ultrasound attenuation (BUA) and determined quantitative ultrasound index (QUI) for phantoms with varying BMD values and varying strontium concentration using two QUS systems: a clinical Sahara® system and an in-house research system with two identical transducers with center frequency of 1 MHz. The two QUS systems were also compared through a Bland-Altman analysis.
Both the clinical system and the research QUS systems showed a strong dependency between BMD and BUA, indicating a potential for QUS to be used as a means of estimating BMD (p = 0.001). SOS was found to have no correlation to BMD (p = 0.546). There was no correlation observed between BUA and increasing bone strontium concentrations for the research (p = 0.749) and clinical (p = 0.609) QUS systems. Similarly, no dependency was observed between the SOS and bone strontium levels up to 2.5 mol/mol [Sr/(Sr+Ca)]% for the research (p = 0.862) and clinical (p = 0.481) QUS systems. No effect on the QUI values was observed with changing strontium levels with either research (p = 0.939) or clinical QUS systems (p = 0.931). A Bland-Altman analysis showed that there was a clear offset in determined QUI values for both systems but they are in agreement with one another.
Bone quality can be assessed through the use of QUS while increasing bone strontium concentration was found to have no effect on QUS-determined quality indices. This study concludes that QUS can potentially be used for the determination of bone quality without introducing biases due to bone strontium levels as is known to be the case with DXA determined aBMD.
目前的金标准双能X线吸收法(DXA)通过测量面积骨密度(aBMD)来测定骨量,如果骨锶水平较高,aBMD会出现正偏差。这对于使用锶基化合物治疗骨质疏松症的人群尤为重要。本研究使用新一代小梁骨模拟体模,研究了定量超声(QUS)测定的骨密度(BMD)以及相关超声测定指标对骨锶含量的依赖性。
开发了一种由羟基磷灰石(HA)和明胶组成的新一代骨模拟体模。这些体模中包含蓖麻油层以创建多层骨模拟体模。使用由锶取代的HA,其锶浓度在0 - 2.5% mol/mol范围内变化的骨矿物质部分来制备这些体模。使用两种QUS系统:临床Sahara®系统和内部研究系统(带有两个中心频率为1MHz的相同换能器),基于不同BMD值和不同锶浓度的体模所测定的声速(SOS)、宽带超声衰减(BUA)和定量超声指数(QUI),评估不同骨锶含量对所测定质量指标的影响。还通过Bland - Altman分析对这两种QUS系统进行了比较。
临床系统和研究QUS系统均显示BMD与BUA之间存在很强的相关性,表明QUS有潜力用作估计BMD的一种方法(p = 0.001)。发现SOS与BMD无相关性(p = 0.546)。对于研究QUS系统(p = 0.749)和临床QUS系统(p = 0.609),未观察到BUA与骨锶浓度增加之间存在相关性。同样,对于研究QUS系统(p = 0.862)和临床QUS系统(p = 0.481),在骨锶水平高达2.5 mol/mol [Sr/(Sr + Ca)]%时,未观察到SOS与骨锶水平之间存在依赖性。无论是研究QUS系统(p = 0.939)还是临床QUS系统(p = 0.931),随着锶水平的变化,未观察到对QUI值有影响。Bland - Altman分析表明,两个系统所测定的QUI值存在明显偏移,但它们彼此一致。
可以通过使用QUS评估骨质量,同时发现增加骨锶浓度对QUS测定的质量指标没有影响。本研究得出结论,QUS有可能用于测定骨质量,而不会像DXA测定aBMD那样因骨锶水平而引入偏差。