Computed Tomography Angiography with a 192-slice Dual-source Computed Tomography System: Improvements in Image Quality and Radiation Dose.

作者信息

Linsen Philip V M, Coenen Adriaan, Lubbers Marisa M, Dijkshoorn Marcel L, Ouhlous Mohamed, Nieman Koen

机构信息

Department of Radiology, Erasmus Medical Center, Rotterdam, The Netherlands.

Department of Radiology, Erasmus Medical Center, Rotterdam, The Netherlands; Department of Cardiology, Erasmus Medical Center, Rotterdam, The Netherlands.

出版信息

J Clin Imaging Sci. 2016 Oct 21;6:44. doi: 10.4103/2156-7514.192840. eCollection 2016.

Abstract

PURPOSE

This study aims to compare image quality, radiation dose, and the influence of the heart rate on image quality of high-pitch spiral coronary computed tomography angiography (CCTA) using 128-slice (second generation) dual-source CT (DSCT) and a 192-slice DSCT (third generation) scanner.

MATERIALS AND METHODS

Two consecutive cohorts of fifty patients underwent CCTA by high-pitch spiral scan mode using 128 or 192-slice DSCT. The 192-slice DSCT system has a more powerful roentgen tube (2 × 120 kW) that allows CCTA acquisition at lower tube voltages, wider longitudinal coverage for faster table speed (732 m/s), and the use of iterative reconstruction. Objective image quality was measured as the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Subjective image quality was evaluated using a Likert scale.

RESULTS

While the effective dose was lower with 192-slice DSCT (1.2 ± 0.5 vs. 0.6 ± 0.3 mSv; < 0.001), the SNR (18.9 ± 4.3 vs. 11.0 ± 2.9; < 0.001) and CNR (23.5 ± 4.8 vs. 14.3 ± 4.1; < 0.001) were superior to 128-slice DSCT. Although patients scanned with 192-slice DSCT had a faster heart rate (59 ± 7 vs. 56 ± 6; = 0.045), subjective image quality was scored higher (4.2 ± 0.8 vs. 3.0 ± 0.7; < 0.001) compared to 128-slice DSCT.

CONCLUSIONS

High-pitch spiral CCTA by 192-slice DSCT provides better image quality, despite a higher average heart rate, at lower radiation doses compared to 128-slice DSCT.

摘要

目的

本研究旨在比较使用128层(第二代)双源CT(DSCT)和192层DSCT(第三代)扫描仪进行高螺距螺旋冠状动脉计算机断层扫描血管造影(CCTA)时的图像质量、辐射剂量以及心率对图像质量的影响。

材料与方法

连续两组各50例患者使用128层或192层DSCT通过高螺距螺旋扫描模式进行CCTA检查。192层DSCT系统具有更强大的X射线管(2×120kW),可在更低的管电压下进行CCTA采集,纵向覆盖范围更广,扫描床速度更快(732m/s),并可使用迭代重建技术。客观图像质量通过信噪比(SNR)和对比噪声比(CNR)进行测量。主观图像质量使用李克特量表进行评估。

结果

192层DSCT的有效剂量更低(1.2±0.5 vs. 0.6±0.3mSv;<0.001),但其SNR(18.9±4.3 vs. 11.0±2.9;<0.001)和CNR(23.5±4.8 vs. 14.3±4.1;<0.001)优于128层DSCT。尽管使用192层DSCT扫描的患者心率更快(59±7 vs. 56±6;=0.045),但与128层DSCT相比,主观图像质量评分更高(4.2±0.8 vs. 3.0±0.7;<0.001)。

结论

与128层DSCT相比,192层DSCT进行的高螺距螺旋CCTA在较低辐射剂量下提供了更好的图像质量,尽管平均心率较高。