Suppr超能文献

一项关于两种血红蛋白仪型号(Hemocue Hb 301和Hb 201+)的方法比较研究,旨在测量柬埔寨柏威夏省女性的血红蛋白浓度并估计贫血患病率。

A method comparison study between two hemoglobinometer models (Hemocue Hb 301 and Hb 201+) to measure hemoglobin concentrations and estimate anemia prevalence among women in Preah Vihear, Cambodia.

作者信息

Rappaport A I, Karakochuk C D, Whitfield K C, Kheang K M, Green T J

机构信息

Food, Nutrition and Health, University of British Columbia, Vancouver, BC, Canada.

NutriFood Cambodia, Phnom Penh, Cambodia.

出版信息

Int J Lab Hematol. 2017 Feb;39(1):95-100. doi: 10.1111/ijlh.12583. Epub 2016 Dec 20.

Abstract

INTRODUCTION

Hemoglobin (Hb) concentration is often measured in global health and nutrition surveys to determine anemia prevalence using a portable hemoglobinometer such as the Hemocue® Hb 201+. More recently, a newer model was released (Hemocue Hb 301) utilizing slightly different methods to measure Hb as compared to the older model. The objective was to measure bias and concordance between Hb concentrations using the Hemocue Hb 301 and Hb 201+ models in a rural field setting.

METHODS

Hemoglobin (Hb) concentration was measured using one finger prick of blood (approximately 10 μL) from 175 Cambodian women (18-49 years) using three Hemocue Hb 201+ and three Hb 301 machines. Bias and concordance were measured and plotted.

RESULTS

Overall, mean ± SD Hb concentration was 116 ± 13 g/L using the Hb 201+ and 118 ± 12 g/L using the Hb 301; and anemia prevalence (Hb < 120 g/L) was 58% (n = 102) and 58% (n = 101), respectively. Overall bias ± SD was 2.0 ± 10.5 g/L and concordance (95% CI) was 0.63 (0.54, 0.72). Despite the 2 g/L bias detected between models, anemia prevalence was very similar in both models.

CONCLUSIONS

The two models measured anemia prevalence comparably in this population of women in rural Cambodia.

摘要

引言

在全球健康与营养调查中,常使用便携式血红蛋白仪(如Hemocue® Hb 201+)测量血红蛋白(Hb)浓度,以确定贫血患病率。最近,一款更新的型号(Hemocue Hb 301)发布,与旧型号相比,其测量Hb的方法略有不同。目的是在农村实地环境中测量使用Hemocue Hb 301和Hb 201+型号测量Hb浓度之间的偏差和一致性。

方法

使用三台Hemocue Hb 201+和三台Hb 301仪器,从175名柬埔寨女性(18 - 49岁)的手指采集一滴血(约10μL)来测量血红蛋白(Hb)浓度。测量并绘制偏差和一致性。

结果

总体而言,使用Hb 201+时Hb浓度的均值±标准差为116±13 g/L,使用Hb 301时为118±12 g/L;贫血患病率(Hb < 120 g/L)分别为58%(n = 102)和58%(n = 101)。总体偏差±标准差为2.0±10.5 g/L,一致性(95%CI)为0.63(0.54, 0.72)。尽管在两种型号之间检测到2 g/L的偏差,但两种型号的贫血患病率非常相似。

结论

在柬埔寨农村的这一女性人群中,这两种型号测量的贫血患病率相当。