老挝儿童中使用 HemoCue 和两种自动化血液分析仪评估血红蛋白的比较。

Comparison of haemoglobin assessments by HemoCue and two automated haematology analysers in young Laotian children.

机构信息

Department of Nutrition, Program in International and Community Nutrition, University of California, Davis, California, USA.

Nutrition and Global Development, Bill & Melinda Gates Foundation, Seattle, Washington, USA.

出版信息

J Clin Pathol. 2018 Jun;71(6):532-538. doi: 10.1136/jclinpath-2017-204786. Epub 2017 Dec 2.

Abstract

BACKGROUND

Haemoglobin (Hb) assessment by Hemocue is used widely for anaemia screening in both adults and children. However, few studies have compared the diagnostic accuracy of Hemocue with an automated haematology analyser in young children.

AIM

To compare Hb concentrations by Hemocue Hb301 and two automated haematology analysers in young children in rural communities of Lao PDR.

METHODS

Capillary blood was collected from 6-month-old to 23-month-old children (n=1487) for determination of Hb concentration by Hemocue Hb301. On the same day, venous blood was collected for complete blood count using one of two haematology analysers (XT-1800i, Sysmex, and BC-3000Plus, Mindray Medical International). In a subsample of children (n=129), venous Hb was also measured by HemoCue Hb301. Agreement between the two methods was estimated using Bland-Altman plots.

RESULTS

Mean capillary Hb by Hemocue was significantly higher than mean venous Hb by haematology analysers combined (108.4±10.3 g/L vs 102.3±13.1 g/L; P<0.001), resulting in a significantly lower anaemia prevalence (Hb <110 g/L) by Hemocue (53.7% vs 73.9%; P<0.001). The Bland-Altman assessment of agreement showed a bias of 6.1 g/L and limits of agreement were -11.5 g/L to 23.7 g/L. Mean venous Hb concentration by Hemocue Hb301 (113.6±14.0 g/L) was significantly higher than mean capillary Hb concentration by Hemocue Hb301 (110.0±10.7; P=0.03 g/L), which in turn was significantly higher than mean venous Hb concentration by the Mindray BC-3000Plus (102.3±17.4 g/L).

CONCLUSION

Capillary and venous Hb concentrations assessed by Hemocue Hb301 showed poor agreement compared with venous Hb by automated haematology analysers, resulting in significantly different anaemia prevalences.

摘要

背景

Hemocue 血红蛋白(Hb)检测被广泛用于成人和儿童的贫血筛查。然而,很少有研究比较 Hemocue 与自动化血液分析仪在幼儿中的诊断准确性。

目的

比较老挝农村社区 6 个月至 23 个月龄儿童的 Hemocue Hb301 与两种自动化血液分析仪的 Hb 浓度。

方法

采集 1487 名 6 个月至 23 个月龄儿童的毛细血管血,用 Hemocue Hb301 检测 Hb 浓度。同日,采集静脉血,用两台血液分析仪(XT-1800i,Sysmex 和 BC-3000Plus,Mindray Medical International)进行全血细胞计数。在儿童亚组(n=129)中,也用 HemoCue Hb301 检测静脉 Hb。采用 Bland-Altman 图估计两种方法的一致性。

结果

Hemocue 毛细血管 Hb 平均值明显高于血液分析仪组合的静脉 Hb 平均值(108.4±10.3 g/L vs 102.3±13.1 g/L;P<0.001),导致 Hemocue 贫血患病率(Hb<110 g/L)显著降低(53.7% vs 73.9%;P<0.001)。一致性的 Bland-Altman 评估显示,存在 6.1 g/L 的偏差,界限为-11.5 g/L 至 23.7 g/L。Hemocue Hb301 静脉 Hb 平均值(113.6±14.0 g/L)明显高于 Hemocue Hb301 毛细血管 Hb 平均值(110.0±10.7 g/L),而后者又明显高于 Mindray BC-3000Plus 的静脉 Hb 平均值(102.3±17.4 g/L)。

结论

Hemocue Hb301 评估的毛细血管和静脉 Hb 浓度与自动化血液分析仪的静脉 Hb 浓度一致性较差,导致贫血患病率显著不同。