Practical approaches to approximating MTF and NPS in CT with an example application to task-based observer studies.

作者信息

Bujila Robert, Fransson Annette, Poludniowski Gavin

机构信息

Medical Radiation Physics and Nuclear Medicine, Karolinska University Hospital, SE-17176 Stockholm, Sweden; Department of Physics, Royal Institute of Technology, SE-106 91 Stockholm, Sweden.

Medical Radiation Physics and Nuclear Medicine, Karolinska University Hospital, SE-17176 Stockholm, Sweden.

出版信息

Phys Med. 2017 Jan;33:16-25. doi: 10.1016/j.ejmp.2016.10.016. Epub 2016 Dec 18.

Abstract

PURPOSE

To investigate two methods of approximating the Modulation Transfer Function (MTF) and Noise Power Spectrum (NPS) in computed tomography (CT) for a range of scan parameters, from limited image acquisitions.

METHODS

The two methods consist of 1) using a linear systems approach to approximate the NPS for different filtered backprojection (FBP) kernels with a filter function derived from the kernel ratio of determined MTFs and 2) using an empirical fitted model to approximate the MTF and NPS. In both cases a scaling function accounts for variations in mAs and kV. The two methods of approximating the MTF/NPS are further investigated by comparing image quality figure of merits (FOM) d' and AUC calculated using approximations of the MTF/NPS and MTF/NPS that have been determined for different mAs/kV levels and reconstruction kernels.

RESULTS

The greatest RMSE for NPS approximated for a range of mAs/kVp/convolution kernels using both methods and compared to determined NPS was 0.05 of the peak value. The RMSE for FOM with the kernel ratio method were at most 0.1 for d' and 0.01 for the AUC. Using the empirical model method, the RMSE for FOM were at most 0.02 for d' and 0.001 for the AUC.

CONCLUSIONS

The two methods proposed in this paper can provide a convenient way of approximating the MTF and NPS for use in, among other things, mathematical observer studies. Both methods require a relatively small number of direct determinations of NPS from scan acquisitions to model the NPS/MTF for arbitrary mAs and kV.

摘要

目的

研究从有限的图像采集中近似计算断层扫描(CT)中的调制传递函数(MTF)和噪声功率谱(NPS)的两种方法,用于一系列扫描参数。

方法

这两种方法包括:1)使用线性系统方法,通过从确定的MTF的核比率导出的滤波函数来近似不同滤波反投影(FBP)核的NPS;2)使用经验拟合模型来近似MTF和NPS。在这两种情况下,缩放函数考虑了毫安秒(mAs)和千伏(kV)的变化。通过比较使用MTF/NPS近似值和已针对不同mAs/kV水平及重建核确定的MTF/NPS计算的图像质量优值(FOM)d'和曲线下面积(AUC),进一步研究这两种近似MTF/NPS的方法。

结果

使用这两种方法针对一系列mAs/kVp/卷积核近似得到的NPS与确定的NPS相比,最大均方根误差(RMSE)为峰值的0.05。核比率法的FOM的RMSE,对于d'最多为0.1,对于AUC最多为0.01。使用经验模型法,FOM的RMSE,对于d'最多为0.02,对于AUC最多为0.001。

结论

本文提出的两种方法可以提供一种方便的方式来近似MTF和NPS,用于数学观察者研究等。两种方法都需要从扫描采集中直接确定相对少量的NPS,以对任意mAs和kV的NPS/MTF进行建模。