Eppeldauer George
National Institute of Standards and Technology, Gaithersburg, MD 20899-0001.
J Res Natl Inst Stand Technol. 1998 Nov-Dec;103(6):615-619. doi: 10.6028/jres.103.040. Epub 1998 Dec 1.
A new method is described to calibrate tristimulus colorimeters for high accuracy color measurements. Instead of traditional lamp standards, modern, high accuracy detector standards are suggested for calibration. After high accuracy absolute spectral response determination of the tristimulus receivers, color (spectral) correction and peak (amplitude) normalization can minimize uncertainties caused by imperfect realizations of the Commission Internationale de l'Eclairage (CIE) color matching functions. As a result of the corrections, stable light sources of different spectral power distributions can be measured with an accuracy dominated by the sub tenths of a percent uncertainty of novel spectral response determinations.
本文描述了一种用于校准三刺激值色度计以进行高精度颜色测量的新方法。建议使用现代高精度探测器标准进行校准,而不是传统的灯标准。在确定三刺激值接收器的高精度绝对光谱响应之后,颜色(光谱)校正和峰值(幅度)归一化可以将由国际照明委员会(CIE)颜色匹配函数的不完美实现所引起的不确定性降至最低。经过校正后,可以以由新型光谱响应测定的百分之零点几的不确定性所主导的精度来测量具有不同光谱功率分布的稳定光源。