用于评估肱骨近端髓内钉抗切割阻力的生物力学测试。

Biomechanical testing to evaluate the cut-through resistance of intramedullary nails for the proximal humerus.

作者信息

Wanzl Maximilian, Foehr Peter, Schreiber Ulrich, Burgkart Rainer H, Lenich Andreas

机构信息

Department of Orthopaedic Sports Surgery, Klinikum rechts der Isar, Technische Universität München, Munich, Germany.

Department of Orthopaedics and Sportsorthopaedics, Klinikum rechts der Isar, Technische Universität München, Munich, Germany.

出版信息

Injury. 2016 Dec;47 Suppl 7:S20-S24. doi: 10.1016/S0020-1383(16)30849-X.

Abstract

Intramedullary nailing for stabilization of proximal humeral fractures is well-established. Complications as part of a cut-through, such as backing out of locking screws, loss of reduction, and perforation of the screws into the glenoid, are equally well-known. The test bench presented in this study enables testing of the cut-through behavior of multiple intramedullary implants on a simulated osteoporotic three-part fracture configuration with three different loading circumstances (A, B and C). In situation A, the glenohumeral dynamic force with progressive loadings entered at an angle of 15° to the humeral shaft. In situation B the force entered at an angle of 35° and in situation C the angle measured 55°. Three different types of nails were tested: the Targon PH with the optimal proximal screw length (T) and with all four proximal screws shortened (Tshort), the Synthes MultiLoc PHN with (S5) and without (S4) the additional calcar screw and, lastly, the PolyAxNail PH, a polyaxial intramedullary nail, in a neutral screw configuration (PAN) and a version with diametrically opposed crossed first and fourth locking screws (PAN10). Significant differences in the three cases were found with the evaluation of the failure load, which represents the cut-through resistance. Case A: Tshort (245.4 ± 18.7 N) - S4 (346.8 ± 18.0 N) (adjusted p = 0.002); Tshort (245.4 ± 18.7 N) - S5 (368.5 ± 12.0 N) (adjusted p = < 0.001); Tshort (245.4 ± 18.7 N) - T (323.5 ± 38.2 N) (p = 0.004); Case B: no significant differences between the study groups (adjusted significance). Case C: PAN (412.5 ± 16.0 N) - S5 (471.5 ± 21.5 N) (adjusted p = 0.007); T (414.0 ± 33.5 N) - S5 (471.5 ± 21.5 N) (adjusted p = 0.008). The optimal screw length has a strong influence on the failure load. Choosing proximal screws that are too short, produces a negative impact on the cut-through resistance. The additional calcar screw of the MultiLoc PHN and the polyaxiality of the PolyAxNail showed a positive effect with regard to the failure load reached.

摘要

髓内钉固定肱骨近端骨折已得到广泛应用。作为穿出相关的并发症,如锁定螺钉退出、复位丢失以及螺钉穿入关节盂等,同样广为人知。本研究中展示的试验台能够在模拟骨质疏松性三部分骨折构型且具有三种不同加载情况(A、B和C)下,测试多种髓内植入物的穿出行为。在情况A中,肱盂动态力以与肱骨干成15°角的方向逐渐加载。在情况B中,力以35°角进入,在情况C中,角度为55°。测试了三种不同类型的髓内钉:具有最佳近端螺钉长度(T)以及所有四个近端螺钉均缩短(Tshort)的Targon PH髓内钉、带有(S5)和不带有(S4)额外距骨螺钉的Synthes MultiLoc PHN髓内钉,以及最后一种PolyAxNail PH髓内钉,一种多轴髓内钉,处于中性螺钉构型(PAN)以及带有直径相对交叉的第一和第四锁定螺钉的版本(PAN10)。通过对代表穿出阻力的失效载荷进行评估,发现三种情况存在显著差异。情况A:Tshort(245.4±18.7 N) - S4(346.8±18.0 N)(调整后p = 0.002);Tshort(245.4±18.7 N) - S5(368.5±12.0 N)(调整后p = < 0.001);Tshort(245.4±18.7 N) - T(323.5±38.2 N)(p = 0.004);情况B:各研究组之间无显著差异(调整后的显著性)。情况C:PAN(412.5±16.0 N) - S5(471.5±21.5 N)(调整后p = 0.007);T(414.0±33.5 N) - S5(471.5±21.5 N)(调整后p = 0.008)。最佳螺钉长度对失效载荷有很大影响。选择过短的近端螺钉会对穿出阻力产生负面影响。MultiLoc PHN的额外距骨螺钉以及PolyAxNail的多轴性在达到的失效载荷方面显示出积极作用。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验