心室噪声和感知不足背景下的房室同步。

Atrioventricular synchrony in the background of ventricular noise and undersensing.

作者信息

Matusik Paweł T, Ząbek Andrzej, Matusik Patrycja S, Małecka Barbara, Lelakowski Jacek

机构信息

Department of Electrocardiology, The John Paul II Hospital, Kraków, Poland.

Jagiellonian University Medical College, Kraków, Poland.

出版信息

Ann Noninvasive Electrocardiol. 2017 Jul;22(4). doi: 10.1111/anec.12425. Epub 2017 Jan 3.

Abstract

The 24-hr electrocardiogram (ECG) interpretation in patients with double-chamber pacemakers may be challenging. The difficulty increases if not well-known pacemaker algorithm and device malfunction coexist. We show atrial synchronization pace (ASP) in a patient with ventricular lead damage. We provide detailed description of electrocardiogram and intracardiac electrogram. ASP may confuse 24-hr ECG monitoring interpretation, especially in patients with ventricular lead dysfunction.

摘要

双腔起搏器患者的24小时心电图(ECG)解读可能具有挑战性。如果不熟悉起搏器算法且同时存在设备故障,困难会增加。我们展示了一名心室导线受损患者的心房同步起搏(ASP)情况。我们提供了心电图和心内电图的详细描述。ASP可能会使24小时ECG监测解读产生混淆,尤其是在心室导线功能障碍的患者中。

相似文献

1
Atrioventricular synchrony in the background of ventricular noise and undersensing.
Ann Noninvasive Electrocardiol. 2017 Jul;22(4). doi: 10.1111/anec.12425. Epub 2017 Jan 3.
3
Switch between AAI and DDD mode pacing-What is the mechanism?
Ann Noninvasive Electrocardiol. 2019 Jul;24(4):e12648. doi: 10.1111/anec.12648. Epub 2019 Mar 21.
4
ECG quiz. What is causing the finding: the pacemaker, patient or the ECG machine?
J Electrocardiol. 2014 Sep-Oct;47(5):752-4. doi: 10.1016/j.jelectrocard.2014.05.003.
5
Unipolar sensing in contemporary pacemakers: using myopotential testing to define optimal sensitivity settings.
J Interv Card Electrophysiol. 1998 Mar;2(1):33-40. doi: 10.1023/a:1009708606035.
6
Diagnosis of atrial undersensing in dual chamber pacemakers: impact of autodiagnostic features.
Pacing Clin Electrophysiol. 1999 Jun;22(6 Pt 1):894-902. doi: 10.1111/j.1540-8159.1999.tb06813.x.
8
Atrial sensing and AV synchrony in single lead VDD pacemakers: a prospective comparison to DDD devices with bipolar atrial leads.
J Cardiovasc Electrophysiol. 1999 Apr;10(4):513-20. doi: 10.1111/j.1540-8167.1999.tb00707.x.
9
Prognostic Significance of Premature Atrial Complexes Burden in Prediction of Long-Term Outcome.
J Am Heart Assoc. 2015 Aug 27;4(9):e002192. doi: 10.1161/JAHA.115.002192.
10
Functional atrial undersensing associated with device algorithm promoting AV conduction.
J Electrocardiol. 2017 Sep-Oct;50(5):696-697. doi: 10.1016/j.jelectrocard.2017.05.009. Epub 2017 May 18.

本文引用的文献

1
Pacing at rates slower than the lower rate limit: function or malfunction?
Pacing Clin Electrophysiol. 2015 Jan;38(1):149-50. doi: 10.1111/pace.12351. Epub 2014 Jan 31.
2
Pacemaker lead displacement: mechanisms and management.
Indian Pacing Electrophysiol J. 2003 Oct 1;3(4):231-8.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验