心脏再同步治疗且依赖起搏器的患者中,因左心室导线故障导致的左心室刺激抑制及左心室T波保护算法

Inhibition of left ventricular stimulation due to left ventricular lead failure and the left ventricular T-wave protection algorithm in patient with cardiac resynchronization therapy and pacemaker dependency.

作者信息

Ząbek Andrzej, Małecka Barbara, Dębski Maciej, Boczar Krzysztof, Lelakowski Jacek

机构信息

Department of Electrocardiology, The John Paul II Hospital, Kraków, Poland.

Institute of Cardiology, Jagiellonian University Medical College, Kraków, Poland.

出版信息

Ann Noninvasive Electrocardiol. 2018 Jan;23(1). doi: 10.1111/anec.12473. Epub 2017 Jun 8.

Abstract

The electrocardiogram (ECG) interpretation in patients with cardiac resynchronization therapy (CRT) is often a perplexing problem. The difficulty in the device evaluation increases in the presence of unfamiliar timing cycles and a lead dysfunction. Authors describe a special function of a Biotronik CRT devices called the left ventricle T-wave protection (LVTP), and demonstrate its behavior in a patient with left ventricular (LV) lead failure. This report shows that sometimes it might be difficult to understand the loss of resynchronization in 12-lead ECG when LVTP feature is on, and a malfunction of left ventricular lead sensing occurs.

摘要

心脏再同步治疗(CRT)患者的心电图(ECG)解读常常是一个令人困惑的问题。在存在不熟悉的定时周期和导联功能障碍的情况下,设备评估的难度会增加。作者描述了百多力CRT设备的一种特殊功能,称为左心室T波保护(LVTP),并展示了其在一名左心室(LV)导联故障患者中的表现。本报告表明,当LVTP功能开启且左心室导联感知出现故障时,有时可能难以理解12导联心电图中再同步的丧失情况。

相似文献

2
Inhibition and restoration of CRT pacing - What is the mechanism?
J Electrocardiol. 2018 May-Jun;51(3):487-489. doi: 10.1016/j.jelectrocard.2018.01.003. Epub 2018 Feb 17.
3
Causes of impaired biventricular pacing in cardiac resynchronization devices with left ventricular sensing.
Pacing Clin Electrophysiol. 2020 Mar;43(3):332-340. doi: 10.1111/pace.13883. Epub 2020 Feb 17.
7
An anterior left ventricular lead position is associated with increased mortality and non-response in cardiac resynchronization therapy.
Int J Cardiol. 2016 Nov 1;222:157-162. doi: 10.1016/j.ijcard.2016.07.235. Epub 2016 Jul 30.
8
T-wave area predicts response to cardiac resynchronization therapy in patients with left bundle branch block.
J Cardiovasc Electrophysiol. 2015 Feb;26(2):176-83. doi: 10.1111/jce.12549. Epub 2014 Oct 20.
9
Cardiac resynchronization therapy by left bundle branch area pacing in patients with heart failure and left bundle branch block.
Heart Rhythm. 2019 Dec;16(12):1783-1790. doi: 10.1016/j.hrthm.2019.09.006. Epub 2019 Sep 9.

本文引用的文献

1
Atrioventricular synchrony in the background of ventricular noise and undersensing.
Ann Noninvasive Electrocardiol. 2017 Jul;22(4). doi: 10.1111/anec.12425. Epub 2017 Jan 3.
2
Implantable Cardiac Defibrillator Lead Failure and Management.
J Am Coll Cardiol. 2016 Mar 22;67(11):1358-68. doi: 10.1016/j.jacc.2015.12.067.
3
Inhibition of Left Ventricular Pacing during Cardiac Resynchronization.
Pacing Clin Electrophysiol. 2015 Aug;38(8):1014-7. doi: 10.1111/pace.12540. Epub 2014 Dec 2.
4
Understanding the timing cycles of a cardiac resynchronization device designed with left ventricular sensing.
Pacing Clin Electrophysiol. 2014 Oct;37(10):1324-37. doi: 10.1111/pace.12496. Epub 2014 Sep 12.
5
Pacemaker rhythm recorded by a cardiac resynchronization device capable of left ventricular sensing.
Pacing Clin Electrophysiol. 2014 Jul;37(7):904-8. doi: 10.1111/pace.12306. Epub 2013 Dec 3.
6
Dissimilar ventricular rhythms: implications for ICD therapy.
Heart Rhythm. 2013 Apr;10(4):510-6. doi: 10.1016/j.hrthm.2012.12.004. Epub 2012 Dec 6.
7
Noise, artifact, and oversensing related inappropriate ICD shock evaluation: ALTITUDE noise study.
Pacing Clin Electrophysiol. 2012 Jul;35(7):863-9. doi: 10.1111/j.1540-8159.2012.03407.x. Epub 2012 Apr 22.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验