Normative Wideband Reflectance, Equivalent Admittance at the Tympanic Membrane, and Acoustic Stapedius Reflex Threshold in Adults.

作者信息

Feeney M Patrick, Keefe Douglas H, Hunter Lisa L, Fitzpatrick Denis F, Garinis Angela C, Putterman Daniel B, McMillan Garnett P

机构信息

1VA RR&D, National Center for Rehabilitative Auditory Research (NCRAR), VA Portland Health Care System, and Oregon Health & Science University, Portland, Oregon, USA; 2Boys Town National Research Hospital, Omaha, Nebraska, USA; and 3Cincinnati Children's Hospital Medical Center, Cincinnati, Ohio, USA.

出版信息

Ear Hear. 2017 May/Jun;38(3):e142-e160. doi: 10.1097/AUD.0000000000000399.

Abstract

OBJECTIVES

Wideband acoustic immittance (WAI) measures such as pressure reflectance, parameterized by absorbance and group delay, equivalent admittance at the tympanic membrane (TM), and acoustic stapedius reflex threshold (ASRT) describe middle ear function across a wide frequency range, compared with traditional tests employing a single frequency. The objective of this study was to obtain normative data using these tests for a group of normal-hearing adults and investigate test-retest reliability using a longitudinal design.

DESIGN

A longitudinal prospective design was used to obtain normative test and retest data on clinical and WAI measures. Subjects were 13 males and 20 females (mean age = 26 years). Inclusion criteria included normal audiometry and clinical immittance. Subjects were tested on two separate visits approximately 1 month apart. Reflectance and equivalent admittance at the TM were measured from 0.25 to 8.0 kHz under three conditions: at ambient pressure in the ear canal and with pressure sweeps from positive to negative pressure (downswept) and negative to positive pressure (upswept). Equivalent admittance at the TM was calculated using admittance measurements at the probe tip that were adjusted using a model of sound transmission in the ear canal and acoustic estimates of ear-canal area and length. Wideband ASRTs were measured at tympanometric peak pressure (TPP) derived from the average TPP of downswept and upswept tympanograms. Descriptive statistics were obtained for all WAI responses, and wideband and clinical ASRTs were compared.

RESULTS

Mean absorbance at ambient pressure and TPP demonstrated a broad band-pass pattern typical of previous studies. Test-retest differences were lower for absorbance at TPP for the downswept method compared with ambient pressure at frequencies between 1.0 and 1.26 kHz. Mean tympanometric peak-to-tail differences for absorbance were greatest around 1.0 to 2.0 kHz and similar for positive and negative tails. Mean group delay at ambient pressure and at TPP were greatest between 0.32 and 0.6 kHz at 200 to 300 μsec, reduced at frequencies between 0.8 and 1.5 kHz, and increased above 1.5 kHz to around 150 μsec. Mean equivalent admittance at the TM had a lower level for the ambient method than at TPP for both sweep directions below 1.2 kHz, but the difference between methods was only statistically significant for the comparison between the ambient method and TPP for the upswept tympanogram. Mean equivalent admittance phase was positive at all frequencies. Test-retest reliability of the equivalent admittance level ranged from 1 to 3 dB at frequencies below 1.0 kHz, but increased to 8 to 9 dB at higher frequencies. The mean wideband ASRT for an ipsilateral broadband noise activator was 12 dB lower than the clinical ASRT, but had poorer reliability.

CONCLUSIONS

Normative data for the WAI test battery revealed minor differences for results at ambient pressure compared with tympanometric methods at TPP for reflectance, group delay, and equivalent admittance level at the TM for subjects with middle ear pressure within ±100 daPa. Test-retest reliability was better for absorbance at TPP for the downswept tympanogram compared with ambient pressure at frequencies around 1.0 kHz. Large peak-to-tail differences in absorbance combined with good reliability at frequencies between about 0.7 and 3.0 kHz suggest that this may be a sensitive frequency range for interpreting absorbance at TPP. The mean wideband ipsilateral ASRT was lower than the clinical ASRT, consistent with previous studies. Results are promising for the use of a wideband test battery to evaluate middle ear function.

摘要

目的

与采用单一频率的传统测试相比,宽带声导抗(WAI)测量,如通过吸光度和群延迟参数化的压力反射率、鼓膜(TM)处的等效导纳以及镫骨肌声反射阈值(ASRT),可在较宽频率范围内描述中耳功能。本研究的目的是使用这些测试为一组听力正常的成年人获取规范数据,并采用纵向设计研究重测信度。

设计

采用纵向前瞻性设计获取关于临床和WAI测量的规范测试及重测数据。受试者为13名男性和20名女性(平均年龄 = 26岁)。纳入标准包括听力正常和临床声导抗正常。受试者在相隔约1个月的两次单独就诊时接受测试。在三种条件下测量TM处的反射率和等效导纳,频率范围为0.25至8.0 kHz:耳道内的环境压力下以及压力从正压到负压(下扫)和从负压到正压(上扫)的压力扫描。TM处的等效导纳通过在探头尖端的导纳测量值计算得出,该测量值使用耳道内声音传播模型以及耳道面积和长度的声学估计值进行了调整。宽带ASRT在从下扫和上扫鼓室图的平均鼓室压峰值(TPP)得出的鼓室压峰值处进行测量。获取所有WAI反应的描述性统计数据,并比较宽带和临床ASRT。

结果

环境压力和TPP下的平均吸光度呈现出先前研究中典型的宽带通模式。与1.0至1.26 kHz频率之间的环境压力相比,下扫法TPP处吸光度的重测差异较小。吸光度的平均鼓室压峰尾差异在1.0至2.0 kHz左右最大,正尾和负尾相似。环境压力和TPP下的平均群延迟在0.32至0.6 kHz之间最大,为200至300微秒,在0.8至1.5 kHz频率之间降低,在1.5 kHz以上增加到约150微秒。在1.2 kHz以下,TM处平均等效导纳在环境方法下的水平低于两个扫描方向的TPP,但方法之间的差异仅在比较上扫鼓室图的环境方法和TPP时具有统计学意义。平均等效导纳相位在所有频率下均为正。等效导纳水平的重测信度在1.0 kHz以下的频率范围为1至3 dB,但在较高频率下增加到8至9 dB。同侧宽带噪声刺激器的平均宽带ASRT比临床ASRT低12 dB,但信度较差。

结论

WAI测试组的规范数据显示,对于中耳压力在±100 daPa范围内的受试者,与TPP处的鼓室图方法相比,环境压力下的反射率、群延迟和TM处等效导纳水平的结果存在细微差异。与1.0 kHz左右频率的环境压力相比,下扫鼓室图TPP处吸光度的重测信度更好。吸光度的大峰尾差异以及在约0.7至3.0 kHz频率之间的良好信度表明,这可能是解释TPP处吸光度的敏感频率范围。平均宽带同侧ASRT低于临床ASRT,与先前研究一致。结果表明使用宽带测试组评估中耳功能具有前景。

相似文献

8
Wideband reflectance in Down syndrome.唐氏综合征中的宽带反射率。
Int J Pediatr Otorhinolaryngol. 2016 Aug;87:164-71. doi: 10.1016/j.ijporl.2016.06.022. Epub 2016 Jun 7.

引用本文的文献

7
Wideband Acoustic Reflex Measurement.宽带声反射测量
Semin Hear. 2023 Mar 14;44(1):84-92. doi: 10.1055/s-0043-1763296. eCollection 2023 Feb.
8
The Rise and Fall of Aural Acoustic Immittance Assessment Tools.耳声导抗评估工具的兴衰
Semin Hear. 2023 Mar 14;44(1):5-16. doi: 10.1055/s-0043-1764139. eCollection 2023 Feb.
10
Brainstem auditory physiology in children with listening difficulties.儿童听力障碍的脑干听觉生理学。
Hear Res. 2023 Mar 1;429:108705. doi: 10.1016/j.heares.2023.108705. Epub 2023 Jan 20.

本文引用的文献

7
Intrasubject variability in power reflectance.功率反射率的个体内变异性。
J Am Acad Audiol. 2014 May;25(5):441-8. doi: 10.3766/jaaa.25.5.3.
8
On reliability.关于可靠性。
Ear Hear. 2014 Sep-Oct;35(5):589-90. doi: 10.1097/AUD.0000000000000042.
10
Wideband acoustic immittance: tympanometric measures.宽频声导抗:鼓室压测量。
Ear Hear. 2013 Jul;34 Suppl 1:65S-71S. doi: 10.1097/AUD.0b013e31829c7250.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验