St Louis R H, Siems W F, Hill H H
Department of Chemistry, Washington State University, Pullman 99164-4630.
J Chromatogr. 1989 Oct 6;479(2):221-31. doi: 10.1016/s0021-9673(01)83338-0.
An ion mobility detector has been designed and constructed for direct axial interfacing with capillary gas chromatography. The principle advantages of this detector were the following: (1) Direct concentric introduction of the capillary column into the ionization region, eliminating peak broadening in the transfer line and improving the efficiency with which neutral molecules were swept from the detector. (2) A variable capillary insertion distance, providing a sensitivity/resolution interplay that could be modified in response to the needs of the assay. (3) An inert gas flow external to the drift cylinder, preventing atmospheric impurities from infiltrating the ultratrace detector. Qualitative and quantitative capabilities of the detector were evaluated using standard preparations of n-hexyl ether.
已设计并构建了一种离子迁移率检测器,用于与毛细管气相色谱进行直接轴向连接。该检测器的主要优点如下:(1) 将毛细管柱直接同心引入电离区域,消除了传输线中的峰展宽,并提高了从检测器中吹扫中性分子的效率。(2) 可变的毛细管插入距离,提供了一种灵敏度/分辨率的相互作用,可根据分析需要进行调整。(3) 在漂移管外部有惰性气体流动,可防止大气杂质渗入超痕量检测器。使用正己基醚标准制剂对该检测器的定性和定量能力进行了评估。