气相色谱后污染对离子迁移率检测的影响。

Effects of contamination on ion mobility detection after gas chromatography.

作者信息

Baim M A, Hill H H

出版信息

J Chromatogr. 1984 Sep 21;299(2):309-19. doi: 10.1016/s0021-9673(01)97847-1.

Abstract

The ion mobility detector is a device that can be used for the selective, ultratrace detection of organic compounds after capillary gas chromatography. It is the only gas chromatographic detector which does not require heteroatomic compounds for selective response, yet concern is often expressed over its quantitative capabilities. Being a secondary ionization device, competitive charge transfer reactions from unseparated compounds or detector gas contamination may decrease the accuracy of measurement. This paper investigates the effects of both electronegative and electro-positive contaminants on the detector's response. In general, it was found that contamination of the detector did affect response but no more severely than in conventional detectors such as the electron-capture detector or the flame ionization detector.

摘要

离子迁移率检测器是一种可用于毛细管气相色谱后对有机化合物进行选择性超痕量检测的装置。它是唯一一种不需要含杂原子化合物进行选择性响应的气相色谱检测器,但人们常常对其定量能力表示担忧。作为一种二次电离装置,未分离化合物的竞争性电荷转移反应或检测器气体污染可能会降低测量的准确性。本文研究了电负性和电正性污染物对检测器响应的影响。一般来说,发现检测器的污染确实会影响响应,但并不比电子捕获检测器或火焰离子化检测器等传统检测器更严重。