Clin Chem. 2017 Mar;63(3):761-769. doi: 10.1373/clinchem.2016.265512. Epub 2017 Jan 18.
The WHO recently released a BK virus (BKV) international standard. This study evaluated the WHO international standard and commercially available BKV standards by quantitative real-time PCR (qPCR) and droplet digital PCR (ddPCR).
WHO, Exact Diagnostics, Acrometrix, and Zeptometrix BKV standards were tested by qPCR and ddPCR. Two preparations of NIST BKV clones were also tested. Nucleic acid was extracted with the Roche MP96 and MPLC, followed by quantification in duplicate. To resolve discrepancies, we sequenced the WHO and NIST materials.
Manufacturers' expected copies/mL were close to WHO IU/mL: linear regression of qPCR data revealed 1.12 Exact copies/IU, 0.76 Acrometrix copies/IU, and 0.70 Zeptometrix copies/IU. For ddPCR, similar concentrations were measured when either the VP1 region or the T region was targeted, and concentrations were almost 2-fold higher when both regions were targeted simultaneously. ddPCR results for the VP1 and T regions were similar for all commercial standards, but targeting the T region of the WHO standard led to a 4-fold lower result than the VP1 region. Next-generation sequencing revealed no primer or probe mismatches. However, large differences in coverage across the WHO standard and junctional reads were observed, indicating subpopulations of the WHO standard with deletions in the T region.
BKV standards showed concordance among providers, but the WHO standard contains subpopulations of viruses with various deletions in the T region. PCR results will vary depending on which region of the WHO standard is targeted.
世界卫生组织(WHO)最近发布了 BK 病毒(BKV)国际标准。本研究通过定量实时 PCR(qPCR)和液滴数字 PCR(ddPCR)对 WHO 国际标准和市售 BKV 标准进行了评估。
通过 qPCR 和 ddPCR 检测了 WHO、Exact Diagnostics、Acrometrix 和 Zeptometrix 的 BKV 标准。还检测了 NIST 的两种 BKV 克隆标准品。使用罗氏 MP96 和 MPLC 提取核酸,然后进行双重定量。为了解决差异,我们对 WHO 和 NIST 材料进行了测序。
制造商预期的拷贝数/mL 与 WHO IU/mL 相近:qPCR 数据的线性回归显示,Exact 的拷贝数/IU 为 1.12,Acrometrix 为 0.76,Zeptometrix 为 0.70。对于 ddPCR,当靶向 VP1 区域或 T 区域时,测量到相似的浓度,当同时靶向两个区域时,浓度几乎增加了两倍。所有商业标准的 ddPCR 结果在 VP1 和 T 区域都相似,但靶向 WHO 标准的 T 区域时,结果比靶向 VP1 区域低 4 倍。下一代测序未发现引物或探针错配。然而,在 WHO 标准和连接读取中观察到覆盖范围的巨大差异,表明 T 区域存在 WHO 标准的亚群,存在缺失。
BKV 标准在供应商之间具有一致性,但 WHO 标准在 T 区域包含具有各种缺失的病毒亚群。PCR 结果将根据 WHO 标准的目标区域而有所不同。