Sun Alexander, Venkatesh A G, Hall Drew A
IEEE Trans Biomed Circuits Syst. 2016 Oct;10(5):945-954. doi: 10.1109/TBCAS.2016.2586504.
This paper describes the design and characterization of a reconfigurable, multi-technique electrochemical biosensor designed for direct integration into smartphone and wearable technologies to enable remote and accurate personal health monitoring. By repurposing components from one mode to the next, the biosensor's potentiostat is able reconfigure itself into three different measurements modes to perform amperometric, potentiometric, and impedance spectroscopic tests all with minimal redundant devices. A [Formula: see text] PCB prototype of the module was developed with discrete components and tested using Google's Project Ara modular smartphone. The amperometric mode has a ±1 nA to [Formula: see text] measurement range. When used to detect pH, the potentiometric mode achieves a resolution of < 0.08 pH units. In impedance measurement mode, the device can measure 50 Ω-10 [Formula: see text] and has been shown to have of phase error. This prototype was used to perform several point-of-care health tracking assays suitable for use with mobile devices: 1) Blood glucose tests were conducted and shown to cover the diagnostic range for Diabetic patients ( ∼ 200 mg/dL). 2) Lactoferrin, a biomarker for urinary tract infections, was detected with a limit of detection of approximately 1 ng/mL. 3) pH tests of sweat were conducted to track dehydration during exercise. 4) EIS was used to determine the concentration of NeutrAvidin via a label-free assay.
本文描述了一种可重构的多技术电化学生物传感器的设计与特性,该传感器专为直接集成到智能手机和可穿戴技术中而设计,以实现远程且准确的个人健康监测。通过将组件从一种模式重新用于另一种模式,该生物传感器的恒电位仪能够将自身重新配置为三种不同的测量模式,以执行安培法、电位法和阻抗光谱测试,且只需最少的冗余设备。使用分立元件开发了该模块的一个[公式:见正文]印刷电路板原型,并使用谷歌的Ara模块化智能手机进行了测试。安培模式的测量范围为±1 nA至[公式:见正文]。在用于检测pH值时,电位模式的分辨率小于0.08 pH单位。在阻抗测量模式下,该设备可测量50 Ω至10[公式:见正文],且已证明存在相位误差。该原型用于执行几种适用于移动设备的即时医疗健康跟踪检测:1) 进行了血糖测试,结果显示覆盖了糖尿病患者的诊断范围(约200 mg/dL)。2) 检测了乳铁蛋白(一种尿路感染的生物标志物),检测限约为1 ng/mL。3) 进行了汗液pH测试,以跟踪运动期间的脱水情况。4) 采用电化学阻抗谱通过无标记检测法测定了中性抗生物素蛋白的浓度。