Metraiah El Hakem Abdelkarim, Regan Helen, Louw Johanna, Kidder Dana
Renal Unit, Aberdeen Royal Infirmary, Aberdeen, UK.
NHS Grampian, Aberdeen, Aberdeen, UK.
BMJ Case Rep. 2017 Jan 23;2017:bcr2016217946. doi: 10.1136/bcr-2016-217946.
Estimation of kidney function by measuring serum creatinine is one the commonest laboratory tests conducted in clinical practice. Enzymatic methods are often used to measure serum creatinine. Clinicians should be aware of the limitations of these methods, such as test interference with paraproteins.We present a case of falsely elevated serum creatinine in a patient referred for renal biopsy. The combination of fluctuating creatinine and normal blood urea level was unusual. Serum protein electrophoresis revealed the presence of an IgM paraprotein. Further investigations confirmed an underlying diagnosis of lymphoplasmacytoid lymphoma. This case highlights how IgM paraprotein can interfere with creatinine estimation by enzymatic assay and the utility of alternative methods of estimating serum creatinine.