Katayama Yoshinori, Enomoto Masahiro, Kikuchi Shin, Takei Atsuko, Ikegami Hitoshi, Minami Hirotaka, Lee Yong Kye
Department of Pediatrics, Takatsuki General Hospital, Takatsuki, Japan.
Department of Pediatrics, Aijinkai Rehabilitation Hospital, Takatsuki, Japan.
Pediatr Int. 2017 Jun;59(6):686-690. doi: 10.1111/ped.13248. Epub 2017 Apr 12.
We tested whether direct transcutaneous bilirubin (TcB) measurement from an area unexposed to phototherapy is reliable for estimation of total serum bilirubin (TSB) in neonates during phototherapy and whether it contributes to reduction in TSB blood sampling in phototherapy decision making.
This was a retrospective observational study of term neonates who received phototherapy in the mother's room. TSB and TcB from the neonate's sternum were measured before and during phototherapy and compared using linear regression analysis and Bland-Altman plot, respectively. Various cut-offs of TcB for estimating TSB during phototherapy at >72 h after birth were analyzed.
There were moderate correlations between TSB and TcB before (r = 0.56) and during (r = 0.47) phototherapy in 125 neonates. The mean difference (TSB-TcB) before and during phototherapy was 1.2 ± 1.7 mg/dL and 1.0 ± 1.7 mg/dL, respectively. The 95% limits of agreement for the difference before and during phototherapy ranged from -2.1 to 4.5 and from -2.3 to 4.3 mg/dL, respectively. For TSB ≤18 mg/dL during phototherapy, a TcB cut-off of 14 mg/dL had a specificity of 1.0; with this method, 43% of the TSB measurements could have been avoided.
Direct measurement of TcB during phototherapy using a bed-type device is a reliable method to estimate TSB in term neonates and would contribute to a reduction in blood sampling. It cannot, however, be used as a substitute for TSB measurement.
我们测试了在光疗期间,从未暴露于光疗区域直接测量经皮胆红素(TcB)对于评估新生儿总血清胆红素(TSB)是否可靠,以及它在光疗决策中是否有助于减少TSB血样采集。
这是一项对在母亲房间接受光疗的足月儿进行的回顾性观察研究。在光疗前和光疗期间测量新生儿胸骨部位的TSB和TcB,并分别使用线性回归分析和Bland-Altman图进行比较。分析了出生后>72小时光疗期间用于估计TSB的不同TcB临界值。
125例新生儿在光疗前(r = 0.56)和光疗期间(r = 0.47),TSB与TcB之间存在中度相关性。光疗前和光疗期间的平均差异(TSB - TcB)分别为1.2±1.7mg/dL和1.0±1.7mg/dL。光疗前和光疗期间差异的95%一致性界限分别为-2.1至4.5和-2.3至4.3mg/dL。对于光疗期间TSB≤18mg/dL,TcB临界值为14mg/dL时特异性为1.0;采用这种方法,43%的TSB测量可以避免。
使用床型设备在光疗期间直接测量TcB是评估足月儿TSB的可靠方法,有助于减少血样采集。然而,它不能替代TSB测量。