在同时进行的PET/MR中,法洛四联症(TOF)与非法洛四联症对PET伪影影响的临床评估:双中心经验

Clinical evaluation of TOF versus non-TOF on PET artifacts in simultaneous PET/MR: a dual centre experience.

作者信息

Ter Voert Edwin E G W, Veit-Haibach Patrick, Ahn Sangtae, Wiesinger Florian, Khalighi M Mehdi, Levin Craig S, Iagaru Andrei H, Zaharchuk Greg, Huellner Martin, Delso Gaspar

机构信息

Department of Nuclear Medicine, University Hospital Zurich, Rämistrasse 100, CH-8091, Zurich, Switzerland.

University of Zurich, Zurich, Switzerland.

出版信息

Eur J Nucl Med Mol Imaging. 2017 Jul;44(7):1223-1233. doi: 10.1007/s00259-017-3619-2. Epub 2017 Jan 26.

Abstract

PURPOSE

Our objective was to determine clinically the value of time-of-flight (TOF) information in reducing PET artifacts and improving PET image quality and accuracy in simultaneous TOF PET/MR scanning.

METHODS

A total 65 patients who underwent a comparative scan in a simultaneous TOF PET/MR scanner were included. TOF and non-TOF PET images were reconstructed, clinically examined, compared and scored. PET imaging artifacts were categorized as large or small implant-related artifacts, as dental implant-related artifacts, and as implant-unrelated artifacts. Differences in image quality, especially those related to (implant) artifacts, were assessed using a scale ranging from 0 (no artifact) to 4 (severe artifact).

RESULTS

A total of 87 image artifacts were found and evaluated. Four patients had large and eight patients small implant-related artifacts, 27 patients had dental implants/fillings, and 48 patients had implant-unrelated artifacts. The average score was 1.14 ± 0.82 for non-TOF PET images and 0.53 ± 0.66 for TOF images (p < 0.01) indicating that artifacts were less noticeable when TOF information was included.

CONCLUSION

Our study indicates that PET image artifacts are significantly mitigated with integration of TOF information in simultaneous PET/MR. The impact is predominantly seen in patients with significant artifacts due to metal implants.

摘要

目的

我们的目标是在临床中确定飞行时间(TOF)信息在减少PET伪影以及提高同时进行的TOF PET/MR扫描中PET图像质量和准确性方面的价值。

方法

纳入了总共65例在同时进行的TOF PET/MR扫描仪中接受对比扫描的患者。重建TOF和非TOF PET图像,进行临床检查、比较并评分。PET成像伪影分为与植入物相关的大或小伪影、与牙种植体相关的伪影以及与植入物无关的伪影。使用从0(无伪影)到4(严重伪影)的量表评估图像质量差异,尤其是与(植入物)伪影相关的差异。

结果

共发现并评估了87个图像伪影。4例患者有与植入物相关的大伪影,8例患者有小伪影,27例患者有牙种植体/填充物,48例患者有与植入物无关的伪影。非TOF PET图像的平均评分为1.14±0.82,TOF图像的平均评分为0.53±0.66(p<0.01),这表明包含TOF信息时伪影不太明显。

结论

我们的研究表明,在同时进行的PET/MR中整合TOF信息可显著减轻PET图像伪影。这种影响在因金属植入物而有明显伪影的患者中最为明显。