Lenters-Westra Erna, English Emma
Clinical Chemistry Department, Isala, Dr. Van Heesweg 2, Zwolle 8025 AB, The Netherlands; European Reference Laboratory for Glycohemoglobin, Isala, Dr. Van Heesweg 2, Zwolle 8025 AB, The Netherlands.
Faculty of Medicine and Health, School of Health Science, University of East Anglia, Norwich Research Park, Norwich NR4 7TJ, UK.
Clin Lab Med. 2017 Mar;37(1):57-71. doi: 10.1016/j.cll.2016.09.006. Epub 2016 Dec 14.
This study uses three unique data sets to show the state of the art of hemoglobin A1c (HbA1c) analyzers in a range of settings and compares their performance against the international guidance set by the International Federation of Clinical Chemistry and Laboratory Medicine task force for HbA1c standardization. The data are used to show the effect of tightening those criteria, and the study serves as a guide to the practical implementation of the sigma-metrics approach in a range of clinical settings.
本研究使用三个独特的数据集来展示糖化血红蛋白(HbA1c)分析仪在一系列环境中的技术水平,并将其性能与国际临床化学和检验医学联合会HbA1c标准化工作组制定的国际指南进行比较。这些数据用于展示收紧这些标准的效果,该研究可作为在一系列临床环境中实际应用西格玛指标方法的指南。